Summary of Safety and Clinical Performance

6.5F Tesio® Catheter

SSCP Document Number: SSCP-010
Revision Number: 6
Revision Date: 5-Sep-25

Important Information

This Summary of Safety and Clinical Performance (SSCP) is intended to provide public access to an updated summary of the main aspects of the safety and clinical performance of the device. This SSCP is not intended to replace the Instructions for Use as the main document to ensure the safe use of the device, nor is it intended to provide diagnostic or therapeutic suggestions to intended users or patients.

Applicable Documents

Document Type Document Title / Number
Design History File (DHF) 96067
‘MDR Documentation’ File Number MDR-010

1. Device Identification and General Information

Device Trade Name(s): 6.5F Tesio® Catheter

Manufacturer Name and Address: Medical Components, Inc. 1499 Delp Drive Harleysville, PA 19438 USA

Manufacturer Single Registration Number (SRN): US-MF-000008230

Basic UDI-DI: 00884908281ND

Medical Device Nomenclature: F900202 – Permanent Hemodialysis Catheter and Kits

Class of Device: III

Date First CE Certificate Issued: 1997

Authorized Representative Name and SRN: European Regulatory Expert Medical Product Service GmbH (MPS) Borngasse 20 35619 Braunfels, Germany SRN: DE-AR-000005009

Notified Body Name and Single Identification Number: BSI Netherlands NB2797

Device Grouping and Variants

The devices in scope of this document are all long-term hemodialysis catheter sets. The device part numbers are organized into variant categories. These devices are distributed as procedure trays, in a configuration inclusive of accessories and adjunctive devices (see section “Accessories intended for use in combination with the Device”).

Variant Devices:

Variant Devices:
Variant Description Part Number(s) Explanation of Multiple Part Numbers
6.5F Tesio 1398 / 1399

Procedure Trays:

Procedure Trays:
Catalog Code Part Number Description
PBFP 1398 / 1399 6.5F Tesio® Catheter Set (Arterial Cuff - 12cm From Tip) (Venous Cuff - 15cm From Tip)

Configurations of Procedure Trays:

Configuration Type Kit Components
6.5F Tesio® Set (2) Catheter (2) 1.3mm OD x 1.0mm ID x 70mm (18GA) Introducer Needle (2) 0.97mm x 70cm (.038) Guidewire J (R 3mm) Tip (2) Advancer (2) Tunneler (2) 2.3mm ID x 14cm (7F) Peelable Introducer (1) Arterial Extension Set (1) Venous Extension Set (2) Clamp (2) End Cap (1) Patient ID Card (1) Patient Information Packet

2. Intended Use of the Device

Intended Purpose: 6.5F Tesio® Catheters are intended for use in pediatric patients who do not have functional permanent vascular access or are not candidates for permanent vascular access for whom central venous vascular access for hemodialysis is deemed necessary based on the direction of a qualified, licensed physician. The catheter is intended to be used under the regular review and assessment of qualified health professionals. This catheter is for single-use only.

Indication(s): The 6.5F Tesio® Catheter is indicated for short-term or long-term use where vascular access is required for 14 days or more for the purpose of hemodialysis.

Target Population(s): 6.5F Tesio® Catheters are intended for use in pediatric patients who do not have functional permanent vascular access or are not candidates for permanent vascular access for whom central venous vascular access for hemodialysis is deemed necessary based on the direction of a qualified, licensed physician.

Contraindications and/or Limitations:

  • Known or suspected allergies to any of the components of the catheter or the kit.
  • This device is contraindicated for patients exhibiting severe, uncontrolled coagulopathy or thrombocytopenia.

3. Device Description

Device Image

Device Name: 6.5F Tesio® Catheter

Description of Device: The 6.5F Tesio® Catheter is a long-term single lumen catheter. Two catheters are inserted into the target vein that are used to remove and return blood through two separate passages (lumens). Priming volume and depth markings are printed on the lumen. A polyester cuff is placed on the catheter’s lumen for tissue ingrowth to anchor the catheter. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. The catheter has been tested at flow rates of up to 250 mL/min.

Device Image

Device Name: 6.5F Tesio® Catheter

Description of Device: The 6.5F Tesio® Catheter is a long-term single lumen catheter. Two catheters are inserted into the target vein that are used to remove and return blood through two separate passages (lumens). Priming volume and depth markings are printed on the lumen. A polyester cuff is placed on the catheter’s lumen for tissue ingrowth to anchor the catheter. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. The catheter has been tested at flow rates of up to 250 mL/min.

Materials / Substances in Contact with Patient Tissue:

The percentages in the table below are based on the weight of the catheter (9.47g).

6.5F Tesio Catheter
Material % Weight (w/w)
Polyvinyl chloride 32.45
Acetal copolymer 40.63
Polyurethane 20.84
Barium sulfate 3.18
Polyethylene terephthalate 1.60
Silicone 1.30

Note:Per the instructions for use, the device is contraindicated for patients with known or suspected allergies to the above materials.

Note:Accessories containing stainless steel may contain up to 4% weight of the CMR substance cobalt.

Information on Medicinal Substances in the Device: N/A

How the Device Achieves its Intended Mode of Action: Hemodialysis catheters are centrally placed access tubes. A typical hemodialysis catheter uses a thin, flexible tube. This catheter has two separate tubes. The tubes go into a large vein. The vein is usually the internal jugular vein. Blood withdraws through one tube of the catheter. The blood flows to the dialysis machine through a separate tubing set. The blood is then processed and filtered. The blood returns to the patient through the second tube. This device is used when dialysis must start at once. Patients may not have a functioning AV fistula or graft. Catheter hemodialysis normally happens on a short-term basis. Long-term access may occur in some cases. For example, when there are problems supporting an AV fistula or graft.

Sterilization Information: Contents sterile and non-pyrogenic in unopened, undamaged package. Sterilized by Ethylene Oxide.

Previous Generations / Variants:

Name of Previous Generation Differences from Current Device
7F Tesio® Catheter N/A

Accessories Intended for Use in Combination with the Device:

Name of Accessory Description of Accessory
Guidewire For general intravascular use to facilitate the selective placement of medical devices in the vessel anatomy.
Guidewire Advancer Aid for introduction of guidewire into target vein.
Introducer Needle Used for the percutaneous introduction of guidewires.
Tunneler Instrument used to create a subcutaneous tunnel
Peelable Introducer Introducers are intended to obtain central venous access to facilitate catheter insertion into the central venous system.
End Cap To protect and keep the catheter clean between treatments
Scalpel A cutting device during surgical, pathology and minor medical procedures
Tegaderm Adhesive wound dressing intended to protect the catheter from contamination when not in use
Syringe Attached to introducer needle to help capture blood return once introducer needle perforates targeted vein, prevent air embolism

Other Devices or Products Intended for Use in Combination with the Device:

Name of Device or Product Description of Device or Product
N/A N/A

4. Risks and Warnings

Residual Risks and Undesirable Effects: As per product IFU (IFU 40775-1BSI), All surgical procedures carry risk. Medcomp has implemented risk management processes to proactively find and mitigate these risks as far as possible without adversely affecting the benefit-risk profile of the device. After mitigation, residual risks and the possibility of adverse events from use of this product remain. Medcomp has determined that all residual risks are acceptable.

Residual Harm Type Possible Adverse Events Associated with Harm
Bleeding Bleeding (May be severe)
Cardiac Event Cardiac Arrhythmia
Embolism Air Embolus
Infection Bacteremia
Perforation Inferior Vena Cava Puncture
Thrombosis Central Venous Thrombosis
Miscellaneous Complications Brachial Plexus Injury
Quantification of Residual Risks
PMS Complaints 01 January 2019 – 31 March 2025 PMCF Events
Units Sold: 687 Units Studied: 0
Patient Residual Harm Category % of Devices % of Devices
Allergic Reaction Not Reported Not Reported
Bleeding Not Reported Not Reported
Cardiac Event Not Reported Not Reported
Embolism Not Reported Not Reported
Infection Not Reported Not Reported
Perforation Not Reported Not Reported
Stenosis Not Reported Not Reported
Tissue Injury Not Reported Not Reported
Thrombosis Not Reported Not Reported

Warnings and Precautions:

All warnings have been reviewed against the risk analysis, PMS, and usability testing to validate consistency between the sources of information. As per product IFUs for all brands (IFU 40775-1BSI), the 6.5F Tesio® Catheter have the following warnings:

  • As per product IFUs for all brands (IFU 40775-1BSI), the 6.5F Tesio® Catheter have the following warnings:
  • Do not insert catheter in thrombosed vessels.
  • Do not advance the guidewire or catheter if unusual resistance is encountered.
  • Do not insert or withdraw the guidewire forcibly from any component. If the guidewire becomes damaged, guidewire and any associated componentry must be removed together.
  • Do not resterilize the catheter or accessories by any method.
  • Contents sterile and non-pyrogenic in unopened, undamaged package. STERILIZED BY ETHYLENE OXIDE
  • Do not re-use catheter or accessories as there may be a failure to adequately clean and decontaminate the device which may lead to contamination, catheter degradation, device fatigue, or endotoxin reaction.
  • Do not use catheter or accessories if package is opened or damaged.
  • Do not use catheter or accessories if any sign of product damage is visible or the use-by date has passed.
  • Do not use sharp instruments near the extension tubing or catheter lumen.
  • Do not use scissors to remove dressing. Precautions listed in the 6.5F Tesio® Catheter IFUs are as follows:
  • Examine catheter lumen and extensions before and after each treatment for damage.
  • To prevent accidents, ensure the security of all caps and bloodline connections prior to and between treatments.
  • Use only Luer Lock (threaded) Connectors with this catheter.
  • In the rare event that a hub or connector separates from any component during insertion or use, take all necessary steps and precautions to prevent blood loss or air embolism and remove the catheter.
  • Before attempting catheter insertion, ensure that you are familiar with the potential complications and their emergency treatment should any of them occur.
  • Repeated overtightening of bloodlines, syringes, and caps will reduce connector life and could lead to potential connector failure.
  • The catheter will be damaged if clamps other than what is provided with this kit are used.
  • Avoid clamping near the Luer Lock and hub of the catheter. Clamping of the tubing repeatedly in the same location may weaken tubing. Additional warnings and cautions listed in the 6.5F Tesio® Catheter IFUs are as follows:
  • Physician discretion is strongly advised when inserting this catheter in patients who are unable to take or hold a deep breath.
  • Patients requiring ventilator support are at increased risk of pneumothorax during subclavian vein cannulation, which may cause complications.
  • Extended use of the subclavian vein may be associated with subclavian vein stenosis.
  • The length of wire inserted is determined by the size of the patient. Monitor patient for signs of arrhythmia throughout this procedure. The patient should be placed on a cardiac monitor during this procedure. Cardiac arrhythmia may result if the guidewire is allowed to pass into the right atrium. The guidewire should be held securely during this procedure.
  • DO NOT grasp and pull the guidewire prior to releasing the J-Straightener. Damage to the guidewire may occur if it is pulled against the restraint of the J-Straightener.
  • DO NOT bend the sheath/dilator during insertion as bending will cause the sheath to prematurely tear. Hold sheath/dilator close to the tip (approximately 3cm from tip) when initially inserting through the skin surface. To progress the sheath/dilator towards the vein, regrasp the sheath/dilator a few centimeters (approximately 5cm) above the original grasp location and push down on the sheath/dilator. Repeat procedure until sheath/dilator is fully inserted.
  • Never leave sheaths in place as indwelling catheters. Damage to the vein will occur.
  • Do not tunnel through muscle.
  • Assure that all air has been aspirated from catheter and the extensions. Failure to do so may result in air embolism.
  • Failure to verify catheter placement may result in serious trauma or fatal complications.
  • Care must be taken when using sharp objects or needles in close proximity to catheter lumen. Contact from sharp objects may cause catheter failure.
  • Only clamp catheter with in-line clamps provided.
  • Extension clamps should only be open for aspiration, flushing, and dialysis treatment.
  • Always review hospital or unit protocol, potential complications and their treatment, warnings, and precautions prior to undertaking any type of mechanical or chemical intervention in response to catheter performance problems.
  • Only a physician familiar with the appropriate techniques should attempt the following procedures.
  • Due to the risk of exposure to HIV (Human Immunodeficiency Virus) or other blood borne pathogens, health care professionals should always use Universal Blood and Body Fluid Precautions in the care of all patients.
  • Do not use excessive force as this may break the catheter. If the catheter does not withdraw from the tunnel after moderate force has been applied, or the catheter is removed without the cuff, the following steps may be taken to remove the catheter.
  • Do not pull distal end of catheter through incision as contamination of wound may occur.

Other Relevant Aspects of Safety: For a period of 01 January 2020 to 31 March 2025 there were 0 complaints for 612 units sold, giving an overall complaint rate of 0.0%. There were no death related events. No events resulted in recalls during the review period.

5. Summary of Clinical Evaluation and Post-Market Clinical Follow-Up (PMCF)

Summary of Clinical Data Related to Subject Device

Specific Case Numbers (Mixed Cohort Case Numbers) Identified and Used for Clinical Performance Evaluation
Product Family Clinical Literature PMCF Data Total Cases User Survey Responses
Clinical Literature 37
PMCF Data 0
Total Cases 37
User Survey Responses 0

Clinical performance was measured using parameters including but not limited to dwell time, catheter insertion outcomes, and adverse event rates. Critical clinical parameters extracted from these studies met standards set forth in the guidelines for the State of the Art. There were no unforeseen adverse events or other high occurrences of adverse events detected in any of the clinical activities. Medcomp® catheters are subjected to, and must pass, simulated use testing intended to replicate use 3 times per week for 12 months as part of device development. The 6.5F Tesio® Catheter passed this testing. Although Medcomp® catheters contain no materials which degrade over time, fully functional catheters may be removed for other reasons, such as intractable infection, change of therapy (such as Renal replacement (transplant) or use of an arterio-venous graft/fistula). Published clinical literature does not always focus on the physical lifetime of a catheter for these reasons. In the case of the 6.5F Tesio® Catheter, 11 catheters had a median 244 day [Range: 1 - 344 days] duration of use that has been found in clinical use reported to date. Based on this information, the 6.5F Tesio® Catheter has a 12-month lifetime; however, the decision to remove and/or replace the catheter should be based on clinical performance and need, and not any predetermined point in time.

Summary of Clinical Data Related to the Equivalent Device

An equivalent device was not used for the device’s clinical evaluation.

Summary of Clinical Data from Pre-Market Investigations (if applicable)

No pre-market clinical devices were used for the device’s clinical evaluation.

Summary of Clinical Data from Other Sources

Source:Summary of Included Literature

Clinical evidence literature searches have found four published literature articles representing 37 6.5F Tesio® device family specific cases. The articles include two prospective studies (Musumeci et al., 2012, Graziano et al., 2014), one retrospective study (Sheth et al., 2001), and one case study (Duarte et al., 2021). Bibliography: Duarte, S.G.G., Alcántara, A., Russo, A., de Sosa, F., Percovich, A.E. (2021). Trans-cells of stent hemodialysis catheter placement in patients with exhausted central venous access Colocación de catéter de hemodiálisis transceldas de stent en paciente con agotamiento de acceso venoso, 73(1), 29 Graziano U, Severino G, Cardone M, Musumeci A, Pecoraro C. Central vascular access by tesio catheters (TC) for chronic hemodialysis in neonates and infants: Beyond the rules. Pediatric Nephrology. 2014;29(9):1760. Musumeci A, Nuzzi F, Graziano U, et al. Use of tesio catheters in infants and children receiving chronic hemodialysis. Pediatric Nephrology. 2012;27(9):1801-1802. Sheth RD, Kale AS, Brewer ED, Brandt ML, Nuchtern JG, Goldstein SL. Successful use of Tesio catheters in pediatric patients receiving chronic hemodialysis. American journal of kidney diseases : the official journal of the National Kidney Foundation. 2001;38(3):553-559.

• Source:PMCF_Medcomp_211

The Medcomp User Survey acquired responses from healthcare personnel familiar with any number of Medcomp’s product offerings. 28 respondents responded that they or their facility have used Medcomp long-term hemodialysis catheters; there were no respondents that have used the 6.5F Tesio® device. There were no differences in mean user sentiments within long-term hemodialysis catheters across State of the Art Performance and Safety Outcome Measures or between device types relating to safety or performance. The following data points were collected from users of Medcomp long-term hemodialysis catheters (n=28):

  • (Mean Likert Scale Response) Catheters function as intended – 4.8 / 5
  • (Mean Likert Scale Response) Packaging allows for aseptic presentation – 4.8 / 5
  • (Mean Likert Scale Response) Benefit outweighs the risk – 4.7 / 5 Dwell Time (n=26) – 167 days (95%CI: 130 – 203)
  • Overall Summary of Clinical Safety and Performance

    Upon review of the 6.5F Tesio® Catheter data across all sources, it is possible to conclude that the benefits of the subject device, which is facilitating hemodialysis in patients in whom other therapies or conservative care are not indicated or desirable as determined by the physician, outweigh the overall and individual risks when the device is used as intended by the manufacturer. It is the manufacturer’s and clinical expert evaluator’s opinion that activities both complete and ongoing are sufficient to support the safety, efficacy, and acceptable benefit/risk profile of the 6.5F Tesio® catheters.

    Overall summary of clinical safety and performance
    Outcome Benefit/Risk Acceptability Criteria Desired Trend Clinical Literature (Subject Device) PMCF Data (Subject Device)
    Performance Outcomes
    Dwell Time Greater than 40 days + 244 days (median) (Summary of Included Literature)
    ND***
    Procedural Outcomes Greater than 93.3% + 100% (Summary of Included Literature)
    ND***
    Safety Outcomes
    Catheter Related Blood Stream Infection (CRBSI) Less than 4.8 incidents of CRBSI per 1,000 catheter days - 1.63/1000 catheter days* (Summary of Included Literature)
    ND***
    Tunnel Infection Rate Less than 2.8 incidents of tunnel infection per 1,000 catheter days - 0.06-2.28/1000 catheter days** (Summary of Included Literature)
    ND***
    Exit Site Infection Rate Less than 3.2 incidents of exit site infection per 1,000 catheter days - 0.06-2.28/1000 catheter days** (Summary of Included Literature)
    ND***
    Catheter Associated Venous Thrombus (CAVT) Less than 3.04 incidents of CAVT per 1,000 catheter days - 0.13/1000 catheter days* (Summary of Included Literature)
    ND***
    *:Event rate is an estimate based on the information in the article **:Includes both Exit Site and Tunnel infection rates ***:ND = No data on the clinical outcome parameter

    On-going or Planned Post-Market Clinical Follow-Up (PMCF)

    Activity Description Reference Timeline
    Multi-center Patient-Level Case Series Collect additional clinical data on the device by acquiring case data healthcare personnel familiar with the device. PMCF_LTHD_241 Q4 2025
    State of the Art Literature Search Identify risks and trends with use of similar devices by reviewing applicable standards, published literature, conference abstracts, guidance documents and recommendations; information relating to the medical condition managed by the device and medical alternatives available for the same target treated population. SAP-HD Q2 2026
    Clinical Evidence Literature Search Identify risks and trends with use of the device by reviewing any clinical data relevant to the device from published literature. LRP-HD Q2 2026
    Global Trial Database Search Identify ongoing clinical trials involving 6.5F Tesio® catheters. N/A Q2 2026

    No emerging risks, complications or unexpected device failures have been detected from PMCF activities.

    6. Possible Therapeutic Alternatives

    The Kidney Disease Outcomes Quality Initiative (KDOQI) 2019 clinical practice guidelines have been used to support the below recommendations for treatments.

    Therapy Benefits Disadvantages Key Risks
    • AV Fistula
    • Permanent vascular access solution
    • Lower complication rate than hemodialysis via catheter
    • Requires time to mature
    • Patients must sometimes self-cannulate
    • Stenosis
    • Thrombosis
    • Aneurysm
    • Pulmonary hypertension
    • Steal Syndrome
    • Septicemia
    • Hemodialysis Catheter
    • Useful for quick vascular access without AV Fistula in place
    • Can be used as a bridge dialysis method between other therapies
    • Not a permanent solution
    • Catheter dysfunction can disrupt regular treatment
    • Benefit is not equal for all patient populations
    • Post-procedural bleeding
    • Infection
    • Thrombosis
    • Decreased blood flow in dysfunctional catheter
    • Cardiovascular events
    • Fibrin sheath formation around catheter
    • Septicemia
    • Peritoneal Dialysis
    • Less restrictive diet than hemodialysis
    • Does not require hospitalization, can be done in any clean place
    • Clearance of impurities is limited by dialysate flow and peritoneal area
    • Peritonitis
    • Septicemia
    • Fluid overload
    • Kidney Transplant
    • Better quality of life compared to HD
    • Lower risk of death compared to HD
    • Fewer dietary restrictions compared to HD
    • Requires a donor which can take time
    • More risky for certain groups (aged, diabetics, etc.)
    • Patient must take rejection medication for life
    • Rejection medication has side effects
    • Thrombosis
    • Hemorrhage
    • Ureteral blockage
    • Infection
    • Organ rejection
    • Death
    • Myocardial infarction
    • Stroke
    • Comprehensive Conservative Care
    • Less imposed symptom burden than dialysis
    • Preserves life satisfaction
    • May aggravate clinical condition
    • Not designed to treat, but to minimize adverse events
    • Treatment may not actually minimize risks associated with CKD
    • AV Fistula (Pediatrics)
    • Preferred pediatric vascular access route
    • Better solute clearance
    • Lower complication rate than hemodialysis with a catheter
    • Lower risk of infection and thrombosis
    • Technical difficulty in fistula/graft creation in children with small vasculature
    • Not suitable for certain patient size
    • High tendency of vasospasm due to small vessels
    • Primary failure and early access thrombosis
    • Hemodialysis Catheter (Pediatrics)
    • Great alternative in rapid onset of kidney failure and short period of time until transplantation
    • Ability to be used in the absence of needle cannulation
    • Decreased risk of high output cardiac failure
    • High infection rates
    • High failure/replacement rate
    • Variable blood flow rates leading to potentially poor clearance
    • Potential complications with significant morbidity and mortality
    • Possible Arrhythmia
    • Permanent damage to central venous system (stenosis/thrombosis) may occur
    • Peritoneal Dialysis (Pediatrics)
    • Most suitable for children due to its almost universal applicability and superior compatibility with lifestyle
    • Long-term success is limited by infectious complications and gradual ultrafiltration failure
    • Catheter exit site and tunnel infection
    • Peritonitis
    • Kidney Transplant (Pediatrics)
    • Enhanced linear growth and potential for remarkable advances in social and intellectual development
    • Graft survival is about 12-15 years in children
    • Increase in the lifetime risk of cancer for pediatric transplant recipients
    • Size – newborns and infants may not be large enough to receive a transplant. Patients need to be around 8-10 kg in size generally
    • Infections, post-transplant lymphoproliferative disorders and malignancy
    • Graft rejection can be difficult to diagnose.

    7. Suggested Profile and Training for Users

    The catheter should be inserted, manipulated, and removed by a qualified, licensed physician or other qualified health care professional under the direction of a physician. In certain circumstances, patients who may be suitable for home hemodialysis may manipulate the external connections of the catheter. As per guidelines stated from the International Society of Hemodialysis, if home dialysis is recommended, each patient will undergo a thorough training in order to obtain optimal results from home dialysis treatments. The objectives of the training program are to (1) provide the appropriate amount of information to ensure that the patient will be able to dialyze safely at home; (2) enable the patient to monitor and manage other elements of his or her chronic kidney disease, such as obtaining samples for lab work and maintaining appropriate nutrition and diet; and (3) help the patient and his or her care partner(s) cope with barriers and fears associated with home HD During training, the patient will also receive technical education on the operations and maintenance of the water treatment system. During training, the ideal nurse trainer-to-patient ratio is typically 1:1. An idealized schedule of training is created, with weekly areas of focus and training objectives. In practice, however, training is individualized to address any identified learning barriers or risks for failure.

    8. Reference to Any Harmonized Standards and Common Specifications (CS) Applied

    Harmonized Standard or CS Revision Title or Description Level of Compliance
    EN ISO 14971 2019+A11:2021 Medical devices. Application of risk management to medical devices Full
    EN ISO 10555-1 2013+A1:2017 Intravascular catheters. Sterile and single-use catheters. General requirements Full
    EN ISO 10555-3 2013 Intravascular catheters. Sterile and single-use catheters. Central venous catheters Full
    EN ISO 11607-1 2020 Packaging for terminally sterilized medical devices. Requirements for materials, sterile barrier systems and packaging systems Full
    EN ISO 11607-2 2020 Packaging for terminally sterilized medical devices. Validation requirements for forming, sealing and assembly processes Full
    MEDDEV 2.7/1 Rev 4 Clinical Evaluation: A Guide for Manufacturers and Notified Bodies Under Directives 93/42/EEC and 90/385/EEC Full
    EN ISO 10993-1 2020 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process Full
    EN ISO 10993-18 2020 Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process Full
    EN ISO 10993-7 2008+ A1:2019 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals — Amendment 1: Applicability of allowable limits for neonates and infants Full
    EN ISO 11135 2014 + A1: 2019 Sterilization of health-care products. Ethylene oxide. Requirements for the development, validation and routine control of a sterilization process for medical devices Full
    ISO 14644-1 2015 Cleanrooms and associated controlled environments — Part 1: Classification of air cleanliness by particle concentration Full
    ISO 14644-2 2015 Cleanrooms and associated controlled environments — Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration Full
    EN 556-1 2001 Sterilization of medical devices. Requirements for medical devices to be designated "STERILE". Requirements for terminally sterilized medical devices Full
    EN ISO 11737-1 2018 + A1: 2021 Sterilization of health care products. Microbiological methods. Determination of a population of microorganisms on products Full
    EN ISO 20417 2021 Medical Devices - Information supplied by the manufacturer Full
    EN ISO 15223-1 2021 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements Full
    ISO 594-1 1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 1: General requirements Full
    ISO 594-2 1998 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 2: Lock Fittings Full
    80369-7 2021 Small-bore connectors for liquids and gases in healthcare applications Part 7: Connectors for intravascular or hypodermic applications Full
    EN 62366-1 2015 + A1: 2020 Medical devices — Part 1: Application of usability engineering to medical devices Full
    ASTM D4332-14 2014 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing Full
    ASTM D4169-16 2016 Standard Practice for Performance Testing of Shipping Containers and Systems Full
    ASTM F2503-20 2020 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment Full
    EN ISO 11070 2014+A1:2018 Sterile single-use intravascular introducers, dilators and guidewires Full
    EN ISO 13485 2016 + A11: 2021 Medical Devices – Quality Management system – Requirements for Regulatory Purposes Full
    ISO/TR 20416 2020 Medical devices — Post-market surveillance for manufacturers Full
    MEDDEV 2.12/2 Rev. 2 GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP STUDIES A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES Full
    MDCG 2020-7 2020 Post-market clinical follow-up (PMCF) Plan Template A guide for manufacturers and notified bodies Full
    MDCG 2020-8 2020 Post-market clinical follow-up (PMCF) Evaluation Report Template A guide for manufacturers and notified bodies Full
    MDCG 2022-9 2022 Summary of safety and clinical performance Full
    MDCG 2018-1 Rev. 4 Guidance on BASIC UDI-DI and changes to UDI-DI Full
    MDCG 2020-6 2020 Clinical evidence needed for medical devices previously CE marked under Directives 93/42/EEC or 90/385/EEC Full
    EN ISO 14155 2020 Clinical investigation of medical devices for human subjects — Good clinical practice Full
    EN ISO 11138-1 2017 Sterilization of health care products — Biological indicators Part 1: General requirements Full
    ISO 11138-2 2017 Sterilization of health care products—Biological indicators—Part 2: Biological indicators for ethylene oxide sterilization processes Full
    ISO 11138-7 2019 Sterilization of health care products. Biological indicators - Guidance for the selection, use and interpretation of results Full
    EN ISO 11140-1 2014 Sterilization of health care products — Chemical indicators Part 1: General requirements Full
    EN ISO/IEC 17025 2017 General requirements for the competence of testing and calibration laboratories Full
    Regulation (EU) 2017/745 2017 Regulation (EU) 2017/745 of the European Parliament and of the Council Full

    Revision History

    Revision Date CR# Author Description of Changes Validated
    1 05OCT2021 26536 RS Implementation of SSCP No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    2 28JUN2022 27030 RS Scheduled Update No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    3 05SEP2022 27250 GM Case numbers sourced from Musumeci et al., 2012 have been corrected to “19” from “21”. This change affects Section 5 (Users / Healthcare Professionals and Patients). Section 8 (Users / Healthcare Professionals) has been updated to match GSPR-010- 02_3. No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    4 10JUL2023 28266 GM Periodic Update; Updated in Accordance with CER-010, Revision D No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    5 01JUL2024 29459 GM Periodic Update; Updating in Accordance with CER-010, Revision E No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    6 05SEP2025 25-0170 GM Periodic Update; Updating in Accordance with CER-010, Revision F No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device

    Version 5.00 of Medical Components, Inc. Template QA-CL-200-1