Important Information
This Summary of Safety and Clinical Performance (SSCP) is intended
to provide public access to an updated summary of the main aspects
of the safety and clinical performance of the device. This SSCP is
not intended to replace the Instructions for Use as the main
document to ensure the safe use of the device, nor is it intended
to provide diagnostic or therapeutic suggestions to intended users
or patients.
Applicable Documents
| Document Type |
Document Title / Number |
| Design History File (DHF) |
96067 |
| ‘MDR Documentation’ File Number |
MDR-010 |
1. Device Identification and General Information
Device Trade Name(s): 6.5F Tesio® Catheter
Manufacturer Name and Address: Medical
Components, Inc. 1499 Delp Drive Harleysville, PA 19438 USA
Manufacturer Single Registration Number (SRN):
US-MF-000008230
Basic UDI-DI: 00884908281ND
Medical Device Nomenclature: F900202 – Permanent
Hemodialysis Catheter and Kits
Class of Device: III
Date First CE Certificate Issued: 1997
Authorized Representative Name and SRN: European
Regulatory Expert Medical Product Service GmbH (MPS) Borngasse 20
35619 Braunfels, Germany SRN: DE-AR-000005009
Notified Body Name and Single Identification Number:
BSI Netherlands NB2797
Device Grouping and Variants
The devices in scope of this document are all long-term
hemodialysis catheter sets. The device part numbers are organized
into variant categories. These devices are distributed as
procedure trays, in a configuration inclusive of accessories and
adjunctive devices (see section “Accessories intended for use in
combination with the Device”).
Variant Devices:
Variant Devices:
| Variant Description |
Part Number(s) |
Explanation of Multiple Part Numbers |
| 6.5F Tesio |
1398 / 1399 |
|
Procedure Trays:
Procedure Trays:
| Catalog Code |
Part Number |
Description |
| PBFP |
1398 / 1399 |
6.5F Tesio® Catheter Set (Arterial Cuff - 12cm From Tip)
(Venous Cuff - 15cm From Tip)
|
Configurations of Procedure Trays:
| Configuration Type |
Kit Components |
| 6.5F Tesio® Set |
(2) Catheter (2) 1.3mm OD x 1.0mm ID x 70mm (18GA)
Introducer Needle (2) 0.97mm x 70cm (.038) Guidewire J (R
3mm) Tip (2) Advancer (2) Tunneler (2) 2.3mm ID x 14cm (7F)
Peelable Introducer (1) Arterial Extension Set (1) Venous
Extension Set (2) Clamp (2) End Cap (1) Patient ID Card (1)
Patient Information Packet
|
2. Intended Use of the Device
Intended Purpose: 6.5F Tesio® Catheters are
intended for use in pediatric patients who do not have functional
permanent vascular access or are not candidates for permanent
vascular access for whom central venous vascular access for
hemodialysis is deemed necessary based on the direction of a
qualified, licensed physician. The catheter is intended to be used
under the regular review and assessment of qualified health
professionals. This catheter is for single-use only.
Indication(s): The 6.5F Tesio® Catheter is
indicated for short-term or long-term use where vascular access is
required for 14 days or more for the purpose of hemodialysis.
Target Population(s): 6.5F Tesio® Catheters are
intended for use in pediatric patients who do not have functional
permanent vascular access or are not candidates for permanent
vascular access for whom central venous vascular access for
hemodialysis is deemed necessary based on the direction of a
qualified, licensed physician.
Contraindications and/or Limitations:
-
Known or suspected allergies to any of the components of the
catheter or the kit.
-
This device is contraindicated for patients exhibiting severe,
uncontrolled coagulopathy or thrombocytopenia.
3. Device Description
Device Name: 6.5F Tesio® Catheter
Description of Device: The 6.5F Tesio® Catheter
is a long-term single lumen catheter. Two catheters are inserted
into the target vein that are used to remove and return blood
through two separate passages (lumens). Priming volume and depth
markings are printed on the lumen. A polyester cuff is placed on
the catheter’s lumen for tissue ingrowth to anchor the catheter.
The catheter incorporates Barium Sulphate to facilitate
visualization under fluoroscopy or Xray. The catheter has been
tested at flow rates of up to 250 mL/min.
Device Name: 6.5F Tesio® Catheter
Description of Device: The 6.5F Tesio® Catheter
is a long-term single lumen catheter. Two catheters are inserted
into the target vein that are used to remove and return blood
through two separate passages (lumens). Priming volume and depth
markings are printed on the lumen. A polyester cuff is placed on
the catheter’s lumen for tissue ingrowth to anchor the catheter.
The catheter incorporates Barium Sulphate to facilitate
visualization under fluoroscopy or Xray. The catheter has been
tested at flow rates of up to 250 mL/min.
Materials / Substances in Contact with Patient Tissue:
The percentages in the table below are based on the weight of the
catheter (9.47g).
6.5F Tesio Catheter
| Material |
% Weight (w/w) |
| Polyvinyl chloride |
32.45 |
| Acetal copolymer |
40.63 |
| Polyurethane |
20.84 |
| Barium sulfate |
3.18 |
| Polyethylene terephthalate |
1.60 |
| Silicone |
1.30 |
Note:Per the instructions for use, the device is contraindicated
for patients with known or suspected allergies to the above
materials.
Note:Accessories containing stainless steel may contain up to 4%
weight of the CMR substance cobalt.
Information on Medicinal Substances in the Device:
N/A
How the Device Achieves its Intended Mode of Action:
Hemodialysis catheters are centrally placed access tubes. A
typical hemodialysis catheter uses a thin, flexible tube. This
catheter has two separate tubes. The tubes go into a large vein.
The vein is usually the internal jugular vein. Blood withdraws
through one tube of the catheter. The blood flows to the dialysis
machine through a separate tubing set. The blood is then processed
and filtered. The blood returns to the patient through the second
tube. This device is used when dialysis must start at once.
Patients may not have a functioning AV fistula or graft. Catheter
hemodialysis normally happens on a short-term basis. Long-term
access may occur in some cases. For example, when there are
problems supporting an AV fistula or graft.
Sterilization Information: Contents sterile and
non-pyrogenic in unopened, undamaged package. Sterilized by
Ethylene Oxide.
Previous Generations / Variants:
| Name of Previous Generation |
Differences from Current Device |
| 7F Tesio® Catheter |
N/A |
Accessories Intended for Use in Combination with the
Device:
| Name of Accessory |
Description of Accessory |
| Guidewire |
For general intravascular use to facilitate the selective
placement of medical devices in the vessel anatomy.
|
| Guidewire Advancer |
Aid for introduction of guidewire into target vein.
|
| Introducer Needle |
Used for the percutaneous introduction of guidewires.
|
| Tunneler |
Instrument used to create a subcutaneous tunnel
|
| Peelable Introducer |
Introducers are intended to obtain central venous access to
facilitate catheter insertion into the central venous
system.
|
| End Cap |
To protect and keep the catheter clean between treatments
|
| Scalpel |
A cutting device during surgical, pathology and minor
medical procedures
|
| Tegaderm |
Adhesive wound dressing intended to protect the catheter
from contamination when not in use
|
| Syringe |
Attached to introducer needle to help capture blood return
once introducer needle perforates targeted vein, prevent air
embolism
|
Other Devices or Products Intended for Use in Combination with
the Device:
| Name of Device or Product |
Description of Device or Product |
| N/A |
N/A |
4. Risks and Warnings
Residual Risks and Undesirable Effects: As per
product IFU (IFU 40775-1BSI), All surgical procedures carry risk.
Medcomp has implemented risk management processes to proactively
find and mitigate these risks as far as possible without adversely
affecting the benefit-risk profile of the device. After
mitigation, residual risks and the possibility of adverse events
from use of this product remain. Medcomp has determined that all
residual risks are acceptable.
| Residual Harm Type |
Possible Adverse Events Associated with Harm
|
| Bleeding |
Bleeding (May be severe)
|
| Cardiac Event |
Cardiac Arrhythmia
|
| Embolism |
Air Embolus
|
| Infection |
Bacteremia
|
| Perforation |
Inferior Vena Cava Puncture
|
| Thrombosis |
Central Venous Thrombosis
|
| Miscellaneous Complications |
Brachial Plexus Injury
|
|
Quantification of Residual Risks
|
|
PMS Complaints 01 January 2019 – 31 March 2025
|
PMCF Events |
|
Units Sold: 687 |
Units Studied: 0 |
|
Patient Residual Harm Category
|
% of Devices |
% of Devices |
| Allergic Reaction |
Not Reported |
Not Reported |
| Bleeding |
Not Reported |
Not Reported |
| Cardiac Event |
Not Reported |
Not Reported |
| Embolism |
Not Reported |
Not Reported |
| Infection |
Not Reported |
Not Reported |
| Perforation |
Not Reported |
Not Reported |
| Stenosis |
Not Reported |
Not Reported |
| Tissue Injury |
Not Reported |
Not Reported |
| Thrombosis |
Not Reported |
Not Reported |
Warnings and Precautions:
All warnings have been reviewed against the risk analysis, PMS,
and usability testing to validate consistency between the sources
of information. As per product IFUs for all brands (IFU
40775-1BSI), the 6.5F Tesio® Catheter have the following warnings:
-
As per product IFUs for all brands (IFU 40775-1BSI), the 6.5F
Tesio® Catheter have the following warnings:
-
Do not insert catheter in thrombosed vessels.
-
Do not advance the guidewire or catheter if unusual resistance
is encountered.
-
Do not insert or withdraw the guidewire forcibly from any
component. If the guidewire becomes damaged, guidewire and any
associated componentry must be removed together.
-
Do not resterilize the catheter or accessories by any method.
-
Contents sterile and non-pyrogenic in unopened, undamaged
package. STERILIZED BY ETHYLENE OXIDE
-
Do not re-use catheter or accessories as there may be a failure
to adequately clean and decontaminate the device which may lead
to contamination, catheter degradation, device fatigue, or
endotoxin reaction.
-
Do not use catheter or accessories if package is opened or
damaged.
-
Do not use catheter or accessories if any sign of product damage
is visible or the use-by date has passed.
-
Do not use sharp instruments near the extension tubing or
catheter lumen.
-
Do not use scissors to remove dressing. Precautions listed in
the 6.5F Tesio® Catheter IFUs are as follows:
-
Examine catheter lumen and extensions before and after each
treatment for damage.
-
To prevent accidents, ensure the security of all caps and
bloodline connections prior to and between treatments.
-
Use only Luer Lock (threaded) Connectors with this catheter.
-
In the rare event that a hub or connector separates from any
component during insertion or use, take all necessary steps and
precautions to prevent blood loss or air embolism and remove the
catheter.
-
Before attempting catheter insertion, ensure that you are
familiar with the potential complications and their emergency
treatment should any of them occur.
-
Repeated overtightening of bloodlines, syringes, and caps will
reduce connector life and could lead to potential connector
failure.
-
The catheter will be damaged if clamps other than what is
provided with this kit are used.
-
Avoid clamping near the Luer Lock and hub of the catheter.
Clamping of the tubing repeatedly in the same location may
weaken tubing. Additional warnings and cautions listed in the
6.5F Tesio® Catheter IFUs are as follows:
-
Physician discretion is strongly advised when inserting this
catheter in patients who are unable to take or hold a deep
breath.
-
Patients requiring ventilator support are at increased risk of
pneumothorax during subclavian vein cannulation, which may cause
complications.
-
Extended use of the subclavian vein may be associated with
subclavian vein stenosis.
-
The length of wire inserted is determined by the size of the
patient. Monitor patient for signs of arrhythmia throughout this
procedure. The patient should be placed on a cardiac monitor
during this procedure. Cardiac arrhythmia may result if the
guidewire is allowed to pass into the right atrium. The
guidewire should be held securely during this procedure.
-
DO NOT grasp and pull the guidewire prior to releasing the
J-Straightener. Damage to the guidewire may occur if it is
pulled against the restraint of the J-Straightener.
-
DO NOT bend the sheath/dilator during insertion as bending will
cause the sheath to prematurely tear. Hold sheath/dilator close
to the tip (approximately 3cm from tip) when initially inserting
through the skin surface. To progress the sheath/dilator towards
the vein, regrasp the sheath/dilator a few centimeters
(approximately 5cm) above the original grasp location and push
down on the sheath/dilator. Repeat procedure until
sheath/dilator is fully inserted.
-
Never leave sheaths in place as indwelling catheters. Damage to
the vein will occur.
- Do not tunnel through muscle.
-
Assure that all air has been aspirated from catheter and the
extensions. Failure to do so may result in air embolism.
-
Failure to verify catheter placement may result in serious
trauma or fatal complications.
-
Care must be taken when using sharp objects or needles in close
proximity to catheter lumen. Contact from sharp objects may
cause catheter failure.
-
Only clamp catheter with in-line clamps provided.
-
Extension clamps should only be open for aspiration, flushing,
and dialysis treatment.
-
Always review hospital or unit protocol, potential complications
and their treatment, warnings, and precautions prior to
undertaking any type of mechanical or chemical intervention in
response to catheter performance problems.
-
Only a physician familiar with the appropriate techniques should
attempt the following procedures.
-
Due to the risk of exposure to HIV (Human Immunodeficiency
Virus) or other blood borne pathogens, health care professionals
should always use Universal Blood and Body Fluid Precautions in
the care of all patients.
-
Do not use excessive force as this may break the catheter. If
the catheter does not withdraw from the tunnel after moderate
force has been applied, or the catheter is removed without the
cuff, the following steps may be taken to remove the catheter.
-
Do not pull distal end of catheter through incision as
contamination of wound may occur.
Other Relevant Aspects of Safety: For a period of
01 January 2020 to 31 March 2025 there were 0 complaints for 612
units sold, giving an overall complaint rate of 0.0%. There were
no death related events. No events resulted in recalls during the
review period.
5. Summary of Clinical Evaluation and Post-Market Clinical
Follow-Up (PMCF)
Summary of Clinical Data Related to Subject Device
Specific Case Numbers (Mixed Cohort Case Numbers) Identified and
Used for Clinical Performance Evaluation
| Product Family |
Clinical Literature |
PMCF Data |
Total Cases |
User Survey Responses |
| Clinical Literature |
37 |
|
|
|
| PMCF Data |
0 |
|
|
|
| Total Cases |
37 |
|
|
|
| User Survey Responses |
0 |
|
|
|
Clinical performance was measured using parameters including but
not limited to dwell time, catheter insertion outcomes, and
adverse event rates. Critical clinical parameters extracted from
these studies met standards set forth in the guidelines for the
State of the Art. There were no unforeseen adverse events or other
high occurrences of adverse events detected in any of the clinical
activities. Medcomp® catheters are subjected to, and must pass,
simulated use testing intended to replicate use 3 times per week
for 12 months as part of device development. The 6.5F Tesio®
Catheter passed this testing. Although Medcomp® catheters contain
no materials which degrade over time, fully functional catheters
may be removed for other reasons, such as intractable infection,
change of therapy (such as Renal replacement (transplant) or use
of an arterio-venous graft/fistula). Published clinical literature
does not always focus on the physical lifetime of a catheter for
these reasons. In the case of the 6.5F Tesio® Catheter, 11
catheters had a median 244 day [Range: 1 - 344 days] duration of
use that has been found in clinical use reported to date. Based on
this information, the 6.5F Tesio® Catheter has a 12-month
lifetime; however, the decision to remove and/or replace the
catheter should be based on clinical performance and need, and not
any predetermined point in time.
Summary of Clinical Data Related to the Equivalent Device
An equivalent device was not used for the device’s clinical
evaluation.
Summary of Clinical Data from Pre-Market Investigations (if
applicable)
No pre-market clinical devices were used for the device’s clinical
evaluation.
Summary of Clinical Data from Other Sources
Source:Summary of Included Literature
Clinical evidence literature searches have found four published
literature articles representing 37 6.5F Tesio® device family
specific cases. The articles include two prospective studies
(Musumeci et al., 2012, Graziano et al., 2014), one retrospective
study (Sheth et al., 2001), and one case study (Duarte et al.,
2021). Bibliography: Duarte, S.G.G., Alcántara, A., Russo, A., de
Sosa, F., Percovich, A.E. (2021). Trans-cells of stent
hemodialysis catheter placement in patients with exhausted central
venous access Colocación de catéter de hemodiálisis transceldas
de stent en paciente con agotamiento de acceso venoso, 73(1), 29
Graziano U, Severino G, Cardone M, Musumeci A, Pecoraro C. Central
vascular access by tesio catheters (TC) for chronic hemodialysis
in neonates and infants: Beyond the rules. Pediatric Nephrology.
2014;29(9):1760. Musumeci A, Nuzzi F, Graziano U, et al. Use of
tesio catheters in infants and children receiving chronic
hemodialysis. Pediatric Nephrology. 2012;27(9):1801-1802. Sheth
RD, Kale AS, Brewer ED, Brandt ML, Nuchtern JG, Goldstein SL.
Successful use of Tesio catheters in pediatric patients receiving
chronic hemodialysis. American journal of kidney diseases : the
official journal of the National Kidney Foundation.
2001;38(3):553-559.
• Source:PMCF_Medcomp_211
The Medcomp User Survey acquired responses from healthcare
personnel familiar with any number of Medcomp’s product offerings.
28 respondents responded that they or their facility have used
Medcomp long-term hemodialysis catheters; there were no
respondents that have used the 6.5F Tesio® device. There were no
differences in mean user sentiments within long-term hemodialysis
catheters across State of the Art Performance and Safety Outcome
Measures or between device types relating to safety or
performance. The following data points were collected from users
of Medcomp long-term hemodialysis catheters (n=28):
(Mean Likert Scale Response) Catheters function as intended – 4.8
/ 5
(Mean Likert Scale Response) Packaging allows for aseptic
presentation – 4.8 / 5
(Mean Likert Scale Response) Benefit outweighs the risk – 4.7 / 5
Dwell Time (n=26) – 167 days (95%CI: 130 – 203)
Overall Summary of Clinical Safety and Performance
Upon review of the 6.5F Tesio® Catheter data across all sources,
it is possible to conclude that the benefits of the subject
device, which is facilitating hemodialysis in patients in whom
other therapies or conservative care are not indicated or
desirable as determined by the physician, outweigh the overall and
individual risks when the device is used as intended by the
manufacturer. It is the manufacturer’s and clinical expert
evaluator’s opinion that activities both complete and ongoing are
sufficient to support the safety, efficacy, and acceptable
benefit/risk profile of the 6.5F Tesio® catheters.
Overall summary of clinical safety and performance
| Outcome |
Benefit/Risk Acceptability Criteria |
Desired Trend |
Clinical Literature (Subject Device) |
PMCF Data (Subject Device) |
| Performance Outcomes |
| Dwell Time |
Greater than 40 days |
+
|
244 days (median) (Summary of Included Literature)
|
ND***
|
| Procedural Outcomes |
Greater than 93.3% |
+
|
100% (Summary of Included Literature)
|
ND***
|
| Safety Outcomes |
|
Catheter Related Blood Stream Infection (CRBSI)
|
Less than 4.8 incidents of CRBSI per 1,000 catheter days
|
-
|
1.63/1000 catheter days* (Summary of Included Literature)
|
ND***
|
| Tunnel Infection Rate |
Less than 2.8 incidents of tunnel infection per 1,000
catheter days
|
-
|
0.06-2.28/1000 catheter days** (Summary of Included
Literature)
|
ND***
|
| Exit Site Infection Rate |
Less than 3.2 incidents of exit site infection per 1,000
catheter days
|
-
|
0.06-2.28/1000 catheter days** (Summary of Included
Literature)
|
ND***
|
|
Catheter Associated Venous Thrombus (CAVT)
|
Less than 3.04 incidents of CAVT per 1,000 catheter days
|
-
|
0.13/1000 catheter days* (Summary of Included Literature)
|
ND***
|
*:Event rate is an estimate based on the information in the
article
**:Includes both Exit Site and Tunnel infection rates
***:ND = No data on the clinical outcome parameter
On-going or Planned Post-Market Clinical Follow-Up (PMCF)
| Activity |
Description |
Reference |
Timeline |
|
Multi-center Patient-Level Case Series
|
Collect additional clinical data on the device by acquiring
case data healthcare personnel familiar with the device.
|
PMCF_LTHD_241 |
Q4 2025 |
| State of the Art Literature Search |
Identify risks and trends with use of similar devices by
reviewing applicable standards, published literature,
conference abstracts, guidance documents and
recommendations; information relating to the medical
condition managed by the device and medical alternatives
available for the same target treated population.
|
SAP-HD |
Q2 2026 |
| Clinical Evidence Literature Search |
Identify risks and trends with use of the device by
reviewing any clinical data relevant to the device from
published literature.
|
LRP-HD |
Q2 2026 |
| Global Trial Database Search |
Identify ongoing clinical trials involving 6.5F Tesio®
catheters.
|
N/A |
Q2 2026 |
No emerging risks, complications or unexpected device failures
have been detected from PMCF activities.
6. Possible Therapeutic Alternatives
The Kidney Disease Outcomes Quality Initiative (KDOQI) 2019
clinical practice guidelines have been used to support the below
recommendations for treatments.
| Therapy |
Benefits |
Disadvantages |
Key Risks |
| • AV Fistula |
-
Permanent vascular access solution
-
Lower complication rate than hemodialysis via catheter
|
- Requires time to mature
-
Patients must sometimes self-cannulate
|
- Stenosis
- Thrombosis
- Aneurysm
- Pulmonary hypertension
- Steal Syndrome
- Septicemia
|
| • Hemodialysis Catheter |
-
Useful for quick vascular access without AV Fistula in
place
-
Can be used as a bridge dialysis method between other
therapies
|
- Not a permanent solution
-
Catheter dysfunction can disrupt regular treatment
-
Benefit is not equal for all patient populations
|
- Post-procedural bleeding
- Infection
- Thrombosis
-
Decreased blood flow in dysfunctional catheter
- Cardiovascular events
-
Fibrin sheath formation around catheter
- Septicemia
|
| • Peritoneal Dialysis |
-
Less restrictive diet than hemodialysis
-
Does not require hospitalization, can be done in any
clean place
|
-
Clearance of impurities is limited by dialysate flow and
peritoneal area
|
- Peritonitis
- Septicemia
- Fluid overload
|
| • Kidney Transplant |
-
Better quality of life compared to HD
-
Lower risk of death compared to HD
-
Fewer dietary restrictions compared to HD
|
-
Requires a donor which can take time
-
More risky for certain groups (aged, diabetics, etc.)
-
Patient must take rejection medication for life
-
Rejection medication has side effects
|
- Thrombosis
- Hemorrhage
- Ureteral blockage
- Infection
- Organ rejection
- Death
- Myocardial infarction
- Stroke
|
| • Comprehensive Conservative Care |
-
Less imposed symptom burden than dialysis
- Preserves life satisfaction
|
- May aggravate clinical condition
-
Not designed to treat, but to minimize adverse events
|
-
Treatment may not actually minimize risks associated
with CKD
|
| • AV Fistula (Pediatrics) |
-
Preferred pediatric vascular access route
- Better solute clearance
-
Lower complication rate than hemodialysis with a
catheter
-
Lower risk of infection and thrombosis
|
-
Technical difficulty in fistula/graft creation in
children with small vasculature
-
Not suitable for certain patient size
|
-
High tendency of vasospasm due to small vessels
-
Primary failure and early access thrombosis
|
| • Hemodialysis Catheter (Pediatrics) |
-
Great alternative in rapid onset of kidney failure and
short period of time until transplantation
-
Ability to be used in the absence of needle cannulation
-
Decreased risk of high output cardiac failure
|
- High infection rates
- High failure/replacement rate
-
Variable blood flow rates leading to potentially poor
clearance
|
-
Potential complications with significant morbidity and
mortality
- Possible Arrhythmia
-
Permanent damage to central venous system
(stenosis/thrombosis) may occur
|
| • Peritoneal Dialysis (Pediatrics) |
-
Most suitable for children due to its almost universal
applicability and superior compatibility with lifestyle
|
-
Long-term success is limited by infectious complications
and gradual ultrafiltration failure
|
-
Catheter exit site and tunnel infection
- Peritonitis
|
| • Kidney Transplant (Pediatrics) |
-
Enhanced linear growth and potential for remarkable
advances in social and intellectual development
-
Graft survival is about 12-15 years in children
|
-
Increase in the lifetime risk of cancer for pediatric
transplant recipients
-
Size – newborns and infants may not be large enough to
receive a transplant. Patients need to be around 8-10 kg
in size generally
|
-
Infections, post-transplant lymphoproliferative
disorders and malignancy
-
Graft rejection can be difficult to diagnose.
|
7. Suggested Profile and Training for Users
The catheter should be inserted, manipulated, and removed by a
qualified, licensed physician or other qualified health care
professional under the direction of a physician. In certain
circumstances, patients who may be suitable for home hemodialysis
may manipulate the external connections of the catheter. As per
guidelines stated from the International Society of Hemodialysis,
if home dialysis is recommended, each patient will undergo a
thorough training in order to obtain optimal results from home
dialysis treatments. The objectives of the training program are to
(1) provide the appropriate amount of information to ensure that
the patient will be able to dialyze safely at home; (2) enable the
patient to monitor and manage other elements of his or her chronic
kidney disease, such as obtaining samples for lab work and
maintaining appropriate nutrition and diet; and (3) help the
patient and his or her care partner(s) cope with barriers and
fears associated with home HD During training, the patient will
also receive technical education on the operations and maintenance
of the water treatment system. During training, the ideal nurse
trainer-to-patient ratio is typically 1:1. An idealized schedule
of training is created, with weekly areas of focus and training
objectives. In practice, however, training is individualized to
address any identified learning barriers or risks for failure.
8. Reference to Any Harmonized Standards and Common Specifications
(CS) Applied
| Harmonized Standard or CS |
Revision |
Title or Description |
Level of Compliance |
| EN ISO 14971 |
2019+A11:2021 |
Medical devices. Application of risk management to medical
devices
|
Full |
| EN ISO 10555-1 |
2013+A1:2017 |
Intravascular catheters. Sterile and single-use catheters.
General requirements
|
Full |
| EN ISO 10555-3 |
2013 |
Intravascular catheters. Sterile and single-use catheters.
Central venous catheters
|
Full |
| EN ISO 11607-1 |
2020 |
Packaging for terminally sterilized medical devices.
Requirements for materials, sterile barrier systems and
packaging systems
|
Full |
| EN ISO 11607-2 |
2020 |
Packaging for terminally sterilized medical devices.
Validation requirements for forming, sealing and assembly
processes
|
Full |
| MEDDEV 2.7/1 |
Rev 4 |
Clinical Evaluation: A Guide for Manufacturers and Notified
Bodies Under Directives 93/42/EEC and 90/385/EEC
|
Full |
| EN ISO 10993-1 |
2020 |
Biological evaluation of medical devices — Part 1:
Evaluation and testing within a risk management process
|
Full |
| EN ISO 10993-18 |
2020 |
Biological evaluation of medical devices — Part 18: Chemical
characterization of medical device materials within a risk
management process
|
Full |
| EN ISO 10993-7 |
2008+ A1:2019 |
Biological evaluation of medical devices — Part 7: Ethylene
oxide sterilization residuals — Amendment 1: Applicability
of allowable limits for neonates and infants
|
Full |
| EN ISO 11135 |
2014 + A1: 2019 |
Sterilization of health-care products. Ethylene oxide.
Requirements for the development, validation and routine
control of a sterilization process for medical devices
|
Full |
| ISO 14644-1 |
2015 |
Cleanrooms and associated controlled environments — Part 1:
Classification of air cleanliness by particle concentration
|
Full |
| ISO 14644-2 |
2015 |
Cleanrooms and associated controlled environments — Part 2:
Monitoring to provide evidence of cleanroom performance
related to air cleanliness by particle concentration
|
Full |
| EN 556-1 |
2001 |
Sterilization of medical devices. Requirements for medical
devices to be designated "STERILE". Requirements for
terminally sterilized medical devices
|
Full |
| EN ISO 11737-1 |
2018 + A1: 2021 |
Sterilization of health care products. Microbiological
methods. Determination of a population of microorganisms on
products
|
Full |
| EN ISO 20417 |
2021 |
Medical Devices - Information supplied by the manufacturer
|
Full |
| EN ISO 15223-1 |
2021 |
Medical devices — Symbols to be used with medical device
labels, labelling and information to be supplied — Part 1:
General requirements
|
Full |
| ISO 594-1 |
1986 |
Conical fittings with a 6 % (Luer) taper for syringes,
needles and certain other medical equipment — Part 1:
General requirements
|
Full |
| ISO 594-2 |
1998 |
Conical fittings with a 6 % (Luer) taper for syringes,
needles and certain other medical equipment — Part 2: Lock
Fittings
|
Full |
| 80369-7 |
2021 |
Small-bore connectors for liquids and gases in healthcare
applications Part 7: Connectors for intravascular or
hypodermic applications
|
Full |
| EN 62366-1 |
2015 + A1: 2020 |
Medical devices — Part 1: Application of usability
engineering to medical devices
|
Full |
| ASTM D4332-14 |
2014 |
Standard Practice for Conditioning Containers, Packages, or
Packaging Components for Testing
|
Full |
| ASTM D4169-16 |
2016 |
Standard Practice for Performance Testing of Shipping
Containers and Systems
|
Full |
| ASTM F2503-20 |
2020 |
Standard Practice for Marking Medical Devices and Other
Items for Safety in the Magnetic Resonance Environment
|
Full |
| EN ISO 11070 |
2014+A1:2018 |
Sterile single-use intravascular introducers, dilators and
guidewires
|
Full |
| EN ISO 13485 |
2016 + A11: 2021 |
Medical Devices – Quality Management system – Requirements
for Regulatory Purposes
|
Full |
| ISO/TR 20416 |
2020 |
Medical devices — Post-market surveillance for manufacturers
|
Full |
| MEDDEV 2.12/2 |
Rev. 2 |
GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP
STUDIES A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES
|
Full |
| MDCG 2020-7 |
2020 |
Post-market clinical follow-up (PMCF) Plan Template A guide
for manufacturers and notified bodies
|
Full |
| MDCG 2020-8 |
2020 |
Post-market clinical follow-up (PMCF) Evaluation Report
Template A guide for manufacturers and notified bodies
|
Full |
| MDCG 2022-9 |
2022 |
Summary of safety and clinical performance
|
Full |
| MDCG 2018-1 |
Rev. 4 |
Guidance on BASIC UDI-DI and changes to UDI-DI
|
Full |
| MDCG 2020-6 |
2020 |
Clinical evidence needed for medical devices previously CE
marked under Directives 93/42/EEC or 90/385/EEC
|
Full |
| EN ISO 14155 |
2020 |
Clinical investigation of medical devices for human subjects
— Good clinical practice
|
Full |
| EN ISO 11138-1 |
2017 |
Sterilization of health care products — Biological
indicators Part 1: General requirements
|
Full |
| ISO 11138-2 |
2017 |
Sterilization of health care products—Biological
indicators—Part 2: Biological indicators for ethylene oxide
sterilization processes
|
Full |
| ISO 11138-7 |
2019 |
Sterilization of health care products. Biological indicators
- Guidance for the selection, use and interpretation of
results
|
Full |
| EN ISO 11140-1 |
2014 |
Sterilization of health care products — Chemical indicators
Part 1: General requirements
|
Full |
| EN ISO/IEC 17025 |
2017 |
General requirements for the competence of testing and
calibration laboratories
|
Full |
| Regulation (EU) 2017/745 |
2017 |
Regulation (EU) 2017/745 of the European Parliament and of
the Council
|
Full |
Revision History
| Revision |
Date |
CR# |
Author |
Description of Changes |
Validated |
| 1 |
05OCT2021 |
26536 |
RS |
Implementation of SSCP |
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 2 |
28JUN2022 |
27030 |
RS |
Scheduled Update |
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 3 |
05SEP2022 |
27250 |
GM |
Case numbers sourced from Musumeci et al., 2012 have been
corrected to “19” from “21”. This change affects Section 5
(Users / Healthcare Professionals and Patients). Section 8
(Users / Healthcare Professionals) has been updated to match
GSPR-010- 02_3.
|
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 4 |
10JUL2023 |
28266 |
GM |
Periodic Update; Updated in Accordance with CER-010,
Revision D
|
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 5 |
01JUL2024 |
29459 |
GM |
Periodic Update; Updating in Accordance with CER-010,
Revision E
|
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 6 |
05SEP2025 |
25-0170 |
GM |
Periodic Update; Updating in Accordance with CER-010,
Revision F
|
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|