Summary of Safety and Clinical Performance

Duo-Flow® Soft-Line® Catheter

SSCP Document Number: SSCP-027
Revision Number: 4
Revision Date: 9/16/2024

Important Information

This Summary of Safety and Clinical Performance (SSCP) is intended to provide public access to an updated summary of the main aspects of the safety and clinical performance of the device. This SSCP is not intended to replace the Instructions for Use as the main document to ensure the safe use of the device, nor is it intended to provide diagnostic or therapeutic suggestions to intended users or patients.

Applicable Documents

Document Type Document Title / Number
Design History File (DHF) 10016, 17007, 17008
‘MDR Documentation’ File Number TD-027

1. Device Identification and General Information

Device Trade Name(s): Duo-Flow® Soft-Line® Catheter

Manufacturer Name and Address: Medical Components, Inc., 1499 Delp Drive, Harleysville, PA 19438 USA

Manufacturer Single Registration Number (SRN): US-MF-000008230

Basic UDI-DI: 00884908301MS

Medical Device Nomenclature: F900201 – Temporary Hemodialysis Catheters and Kits

Class of Device: III

Date First CE Certificate Issued: 1-Mar

Authorized Representative Name and SRN: Gerhard Frömel European Regulatory Expert Medical Product Service GmbH (MPS) Borngasse 20 35619 Braunfels, Germany SRN: DE-AR-000005009

Notified Body Name and Single Identification Number: BSI Group The Netherlands B.V. NB2797

Device Grouping and Variants

The devices in scope of this document are all short-term hemodialysis catheter sets. The device part numbers are organized into variant categories. These devices are distributed as procedure trays, in various configurations inclusive of accessories and adjunctive devices (see section “Accessories intended for use in combination with the Device”).

Variant Devices:

Variant Devices:
Variant Description Part Number(s) Explanation of Multiple Part Numbers
11.5F x 12cm Straight Duo-Flow Soft-Line 1348G
11.5F x 15cm Pre-Curved Duo-Flow Soft-Line 1388G
11.5F x 15cm Straight Duo-Flow Soft-Line 1346G
11.5F x 20cm Pre-Curved Duo-Flow Soft-Line 1389G
11.5F x 20cm Straight Duo-Flow Soft-Line 1347G
11.5F x 24cm Straight Duo-Flow Soft-Line 1364G
7F x 10cm Straight Duo-Flow Soft-Line 1352G-10
7F x 7cm Straight Duo-Flow Soft-Line 1352G-7
9F x 12cm Straight Duo-Flow Soft-Line 1349
9F x 15cm Straight Duo-Flow Soft-Line 1350
9F x 20cm Straight Duo-Flow Soft-Line 1351

Procedure Trays:

Procedure Trays:
Catalog Code Part Number Description
DJST710 1352G-10 7F X 10cm Duo-Jet® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
DL7/10 1352G-10 7F X 10cm Nikkiso Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T114ME 1348G 11.5F X 12cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T116IJS-2E. 1388G 11.5F X 15cm Duo-Flow® Soft-Line® Double Lumen Pre-Curved Hemodialysis Catheter W/ Dual Suture Wing Basic Set
T116ME 1346G 11.5F X 15cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T118IJS-2E. 1389G 11.5F X 20cm Duo-Flow® Soft-Line® Double Lumen Pre-Curved Hemodialysis Catheter W/ Dual Suture Wing Basic Set
T118ME 1347G 11.5F X 20cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T119M 1364G 11.5F X 24cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T73M 1352G-7 7F X 7cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T74M 1352G-10 7F X 10cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T94M 1349 9F X 12cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T96M 1350 9F X 15cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set
T98M 1351 9F X 20cm Duo-Flow® Soft-Line® Double Lumen Hemodialysis Catheter Basic Set

Configurations of Procedure Trays:

Configuration Type Kit Components
Duo-Flow® Soft-Line® Basic Set (1) Catheter (1) Guidewire (1) Guidewire Advancer (1) Needle (1) Scalpel (1) Dilator (2) End Cap

2. Intended Use of the Device

Intended Purpose: The Duo-Flow® Soft-Line® Catheters are intended for use in adult patients with Acute Kidney Injury (AKI) or Chronic Kidney Disease (CKD) for whom immediate central venous vascular access for short term hemodialysis is deemed necessary based on the direction of a qualified, licensed physician. The catheter is intended to be used under the regular review and assessment of qualified health professionals. This catheter is for Single Use Only.

Indication(s): The Duo-Flow® Soft-Line® Catheter is indicated for short-term use where vascular access is required for less than 14 days for the purpose of hemodialysis.

Target Population(s): Duo-Flow® Soft-Line® catheters are intended for use in adult patients with Acute Kidney Injury (AKI) or Chronic Kidney Disease (CKD) for whom immediate central venous vascular access for short-term hemodialysis is deemed necessary based on the direction of a qualified, licensed physician. The catheter is not intended for use in pediatric patients.

Contraindications and/or Limitations:

  • Known or suspected allergies to any of the components of the catheter or the kit.
  • This device is contraindicated for patients exhibiting severe, uncontrolled coagulopathy or thrombocytopenia.

3. Device Description

Device Image

Device Name: Duo-Flow® Soft-Line® Catheter

Description of Device: Duo-Flow® Soft-Line®: The Duo-Flow® Soft-Line® Catheter removes and returns blood through two segregated lumen passages. Each lumen is connected to an extension line with color-coded female luer connectors. The transition between lumen and extension is housed within a molded hub. Both arterial and venous lumens contain side-holes. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. The catheter is available with a straight or pre-curved lumen in a variety of French sizes and lengths to accommodate physician preference and clinical needs.

Device Image

Device Name: Duo-Jet® Soft-Line® Catheter; Nikkiso Soft-Line® Catheter

Description of Device: Duo-Jet® Soft-Line®: The Duo-Jet® Soft-Line® Catheter removes and returns blood through two segregated lumen passages. Each lumen is connected to an extension line with color-coded female luer connectors. The transition between lumen and extension is housed within a molded hub. Both arterial and venous lumens contain side-holes. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. Nikkiso Soft-Line®: The Nikkiso Soft-Line® Catheter removes and returns blood through two segregated lumen passages. Each lumen is connected to an extension line with color-coded female luer connectors. The transition between lumen and extension is housed within a molded hub. Both arterial and venous lumens contain side-holes. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray.

Materials / Substances in Contact with Patient Tissue:

The percentage ranges in the table below are based on the weights of the 7F x 7cm catheter (7.48g) and the 11.5F x 20cm catheter (9.94g).

Duo-Flow® Soft-Line®
Material % Weight (w/w)
Polyurethane 37.75 - 46.38
Acetal co-polymer 19.28 – 25.63
Polyvinyl chloride 19.23 – 25.56
Acrylonitrile Butadiene Styrene 7.81 - 10.38
Barium sulfate 0.68 - 5.01
Vythene 0 - 2.28

Note:Per the instructions for use, the device is contraindicated for patients with known or suspected allergies to the above materials.

Information on Medicinal Substances in the Device: N/A

How the Device Achieves its Intended Mode of Action: Hemodialysis catheters are centrally placed access tubes. A typical hemodialysis catheter uses a thin, flexible tube. The tube has two openings. The tube goes into a large vein. The vein is usually the internal jugular vein. Blood withdraws through one lumen of the catheter. The blood flows to the dialysis machine through a separate tubing set. The blood is then processed and filtered. The blood returns to the patient through the second lumen. This device is used when dialysis must start at once. Patients may not have a functioning AV fistula or graft. Catheter hemodialysis normally happens on a short-term basis.

Sterilization Information: Contents sterile and non-pyrogenic in unopened, undamaged package. Sterilized by Ethylene Oxide.

Previous Generations / Variants:

Name of Previous Generation Differences from Current Device
N/A N/A

Accessories Intended for Use in Combination with the Device:

Name of Accessory Description of Accessory
Guidewire For general intravascular use to facilitate the selective placement of medical devices in the vessel anatomy.
Guidewire Advancer Aid for introduction of guidewire into target vein.
Introducer Needle Used for the percutaneous introduction of guidewires.
Scalpel A cutting device during surgical, pathology and minor medical procedures
Dilator Designed for percutaneous entry into a vessel in order to enlarge the opening of the vessel for the placement of a catheter in a vein.
End Cap To keep clean and protect catheter luer between treatments.

Other Devices or Products Intended for Use in Combination with the Device:

Name of Device or Product Description of Device or Product
Syringe Attached to introducer needle to help capture blood return once introducer needle perforates targeted vein, prevent air embolism

4. Risks and Warnings

Residual Risks and Undesirable Effects: All surgical procedures carry risk. Medcomp® has implemented risk management processes to proactively find and mitigate these risks as far as possible without adversely affecting the benefit-risk profile of the device. After mitigation, residual risks and the possibility of adverse events from use of this product remain. Medcomp® has determined that all residual risks are acceptable when considered with respect to the expected clinical benefits of the Duo-Flow® Soft-Line® Catheter and the benefits of other similar hemodialysis devices.

Residual Harm Type Possible Adverse Events Associated with Harm
Allergic Reaction Allergic Reaction
Bleeding Bleeding (May be severe)
Cardiac Event Cardiac Arrhythmia
Embolism Air Embolus
Infection Bacteremia
Perforation Inferior Vena Cava Puncture
Stenosis Venous Stenosis
Tissue Injury Brachial Plexus Injury
Thrombosis Central Venous Thrombosis
Miscellaneous Complications Catheter Dysfunction
Quantification of Residual Risks
PMS Complaints 01 January 2018 – 31 December 2023 PMCF Events
Units Sold: 28,628 Units Studied: 0
Patient Residual Harm Category % of Devices % of Devices
Allergic Reaction Not Reported Not Reported
Bleeding 0.003% Not Reported
Cardiac Event Not Reported Not Reported
Embolism Not Reported Not Reported
Infection Not Reported Not Reported
Perforation Not Reported Not Reported
Stenosis Not Reported Not Reported
Tissue Injury Not Reported Not Reported
Thrombosis Not Reported Not Reported

Warnings and Precautions:

Warnings listed for the Duo-Flow® Soft-Line® Catheter are as follows:

  • Do not insert catheter in thrombosed vessels.
  • Do not advance the guidewire or catheter if unusual resistance is encountered.
  • Do not insert or withdraw the guidewire forcibly from any component. If the guidewire becomes damaged, guidewire and any associated componentry must be removed together.
  • Do not resterilize the catheter or accessories by any method.
  • Contents sterile and non-pyrogenic in unopened, undamaged package. STERILIZED BY ETHYLENE OXIDE
  • Do not re-use catheter or accessories as there may be a failure to adequately clean and decontaminate the device which may lead to contamination, catheter degradation, device fatigue, or endotoxin reaction.
  • Do not use catheter or accessories if package is opened or damaged.
  • Do not use catheter or accessories if any sign of product damage is visible or the use-by date has passed.
  • Do not use sharp instruments near the extension tubing or catheter lumen.
  • Do not use scissors to remove dressing. Precautions listed for Duo-Flow® Soft-Line® Catheter are as follows:
  • Examine catheter lumen and extensions before and after each treatment for damage.
  • To prevent accidents, ensure the security of all caps and bloodline connections prior to and
  • between treatments.
  • Use only Luer Lock (threaded) Connectors with this catheter.
  • In the rare event that a hub or connector separates from any component during insertion or use, take all necessary steps and precautions to prevent blood loss or air embolism and remove the catheter.
  • Before attempting catheter insertion, ensure that you are familiar with the potential complications and their emergency treatment should any of them occur.
  • Repeated overtightening of bloodlines, syringes, and caps will reduce connector life and could lead to potential connector failure.
  • The catheter will be damaged if clamps other than what is provided with this kit are used.
  • Avoid clamping near the Luer Lock and hub of the catheter. Clamping of the tubing repeatedly in the same location may weaken tubing.

Other Relevant Aspects of Safety: For a period of 01 January 2019 to 31 December 2023 there were 12 complaints for 24,408 units sold, giving an overall complaint rate of 0.049%. There were no death-related events. No events resulted in recalls during the review period.

5. Summary of Clinical Evaluation and Post-Market Clinical Follow-Up (PMCF)

Summary of Clinical Data Related to Subject Device

Specific Case Numbers (Mixed Cohort Case Numbers) Identified and Used for Clinical Performance Evaluation
Product Family Clinical Literature PMCF Data Total Cases User Survey Responses
70 (& 88 Mixed Cohort Cases) 0 70 (& 88 Mixed Cohort Cases) 1

Clinical performance was measured using parameters including but not limited to dwell time, catheter insertion outcomes, and adverse event rates. Critical clinical parameters extracted from these studies met standards set forth in the guidelines for the State of the Art. There were no unforeseen adverse events or other high occurrences of adverse events detected in any of the clinical activities. Medcomp® STHD catheters are subjected to, and must pass, simulated use testing intended to replicate 30 days use as part of device development. The Duo-Flow® Soft-Line® Catheter passed this testing. Clinical guidelines recommend to limit the use of temporary, noncuffed, nontunled dialysis catheters to a maximum of 2 weeks (KDOQI 2019), however, duration of use of these catheters has varied in available clinical evidence identified by the manufacturer to date. Although Medcomp® catheters materials contain non-degradable polymers, fully functional catheters may be removed for other reasons, such as intractable infection or change of therapy. Published clinical literature does not always focus on the physical lifetime of a catheter for these reasons. In the case of the Duo-Flow® Soft-Line® Catheter, 69 catheters had a mean dwell time of 36.4 days [95% CI: 0 – 73.6 days] duration of use that has been found in clinical use reported to date. Based on this information, the Duo-Flow® Soft-Line® catheters have a 30 day lifetime; however, the decision to remove and/or replace the catheter should be based on clinical performance and need, and not any predetermined point in time.

Summary of Clinical Data Related to the Equivalent Device

  • Clinical evidence from published literature and PMCF activities has been generated specific to known and unknown variants of the subject device. The equivalency rationale in the updated clinical evaluation report will demonstrate that the clinical evidence available for these variants is representative of the range of device variants in the device family. Variants Relying on Equivalence:
  • Specific Duo-Flow® Soft-Line® Variants Variants Contributing Clinical Data:
  • Duo-Flow® Soft-Line® (Unknown Variant) There are no clinical or biological differences between variants within the subject device family, and the potential impact of the technical differences will be rationalized in the updated clinical evaluation report.
  • Summary of Clinical Data from Pre-Market Investigations (if applicable)

    No pre-market clinical devices were used for the device’s clinical evaluation.

    Summary of Clinical Data from Other Sources

    Source:Summary of Published Literature

    Clinical evidence literature searches have found three published literature articles representing 70 Duo-Flow® Soft-Line® device family specific cases, and an additional 88 mixed cohort cases inclusive of Duo-Flow® Soft-Line® devices. The articles included two retrospective studies (Amira et al. and Park et al.) and one case report (Fuentes et al.). Bibliography: Amira CO, Bello BT, Braimoh RW. A study of outcome and complications associated with temporary hemodialysis catheters in a Nigerian dialysis unit. Saudi journal of kidney diseases and transplantation: an official publication of the Saudi Center for Organ Transplantation, Saudi Arabia. 2016;27(3):569-75. Fuentes, A. D., Rubio, G. T., Acuña, C. A., Rubio, F. D., Milic, F. B., & Troncoso, P. C. (2023). Near-fatal cocaine intoxication in an infant with thrombotic microangiopathy associated with multiple organ failure. Revista Paulista de Pediatria, 42, e2022159. Park HS, Choi J, Kim HW et al. Exchange over the guidewire from non-tunneled to tunneled hemodialysis catheters can be performed without patency loss. The Journal of Vascular Access. 2018;19(3):252-7.

    • Source:PMCF_Medcomp_211

    The Medcomp User Survey acquired responses from healthcare personnel familiar with any number of Medcomp’s product offerings. 20 respondents responded that they or their facility have used Medcomp short-term hemodialysis catheters, with 1 of those respondents using the Duo-Flow® Soft-Line® device. There were no differences in mean user sentiments within short-term hemodialysis catheters across State of the Art Performance and Safety Outcome Measures or between device types relating to safety or performance. The following data points were collected from users of Medcomp short-term hemodialysis catheters (n=20):

  • (Mean Likert Scale Response) Catheters function as intended – 4.8 / 5
  • (Mean Likert Scale Response) Packaging allows for aseptic presentation – 4.9 / 5
  • (Mean Likert Scale Response) Benefit outweighs the risk – 4.7 / 5
  • Dwell Time (n=19) – 15.74 days (95%CI: 6.3 – 25.1) The following data points were collected from users of Medcomp Duo-Flow® Soft-Line® catheters (n=1):
  • (Mean Likert Scale Response) Catheters function as intended – 5 / 5
  • (Mean Likert Scale Response) Packaging allows for aseptic presentation – 5 / 5
  • (Mean Likert Scale Response) Benefit outweighs the risk – 5 / 5
  • Dwell Time (n=1) – 14 days
  • Overall Summary of Clinical Safety and Performance

    Upon review of the Duo-Flow® Soft-Line® catheter data across all sources, it is possible to conclude that the benefits of the subject device outweigh the overall and individual risks when the device is used as intended by the manufacturer. It is the manufacturer’s and clinical expert evaluator’s opinion that activities both complete and ongoing are sufficient to support the safety, efficacy, and acceptable benefit/risk profile of the subject devices.

    Overall summary of clinical safety and performance
    Outcome Benefit/Risk Acceptability Criteria Desired Trend Clinical Literature (Subject Device) PMCF Data (Subject Device)
    Performance Outcomes
    Dwell Time Greater than 8 days + 36.4 days (Summary of Published Literature)
    14 days (PMCF_Medcomp_211) Likert Scale Response 5 / 5 (PMCF_Medcomp_211)**
    Procedural Outcomes Greater than 95% + No insertion complications reported. (Summary of Published Literature)
    Likert Scale Response 5 / 5 (PMCF_Medcomp_211)**
    Safety Outcomes
    Catheter Related Blood Stream Infection (CRBSI) Less than 7.8 incidents of CRBSI per 1,000 catheter days - 0.25 – 1.73 per 1,000 catheter days (Summary of Published Literature)
    Likert Scale Response 5 / 5 (PMCF_Medcomp_211)**
    Exit Site Infection Rate Less than 3.5 incidents of exit site infection per 1,000 catheter days - ND*
    Likert Scale Response 5 / 5 (PMCF_Medcomp_211)**
    Catheter Associated Venous Thrombus (CAVT) Less than 11.4 incidents of CAVT per 1,000 catheter days - 9.88 per 1,000 catheter days (Summary of Published Literature)
    Likert Scale Response 5 / 5 (PMCF_Medcomp_211)**
    *ND indicates no data on the clinical data parameter: **PMCF_Medcomp_211 asked respondents, if they agreed on a scale of 1 -5, that their experience in relation to each outcome was the same or better than the benefit/risk acceptability criteria.:

    On-going or Planned Post-Market Clinical Follow-Up (PMCF)

    Activity Description Reference Timeline
    Multicenter Patient-Level Case Series Collect additional clinical data on the device PMCF_STHD_241 Q4 2025
    State of the Art Literature Search Identify risks and trends with use of dialysis catheters SAP-HD Q1 2025
    Clinical Evidence Literature Search Identify risks and trends with use of the device LRP-STHD Q3 2025
    Global Trial Database Search Identify ongoing clinical trials involving Medcomp® catheters N/A Q3 2025

    No emerging risks, complications or unexpected device failures have been detected from PMCF activities.

    6. Possible Therapeutic Alternatives

    The Kidney Disease Outcomes Quality Initiative (KDOQI) 2019 clinical practice guidelines have been used to support the below recommendations for treatments.

    Therapy Benefits Disadvantages Key Risks
    • AV Fistula
    • Permanent vascular access solution
    • Lower complication rate than hemodialysis via catheter
    • Requires time to mature
    • Patients must sometimes self-cannulate
    • Stenosis
    • Thrombosis
    • Aneurysm
    • Pulmonary hypertension
    • Steal Syndrome
    • Septicemia
    • Hemodialysis Catheter
    • Useful for quick vascular access without AV Fistula in place
    • Can be used as a bridge dialysis method between other therapies
    • Not a permanent solution
    • Catheter dysfunction can disrupt regular treatment
    • Benefit is not equal for all patient populations
    • Post-procedural bleeding
    • Infection
    • Thrombosis
    • Decreased blood flow in dysfunctional catheter
    • Cardiovascular events
    • Fibrin sheath formation around catheter
    • Septicemia
    • Peritoneal Dialysis
    • Less restrictive diet than hemodialysis
    • Does not require hospitalization, can be done in any clean place
    • Clearance of impurities is limited by dialysate flow and peritoneal area
    • Peritonitis
    • Septicemia
    • Fluid overload
    • Kidney Transplant
    • Better quality of life compared to HD
    • Lower risk of death compared to HD
    • Fewer dietary restrictions compared to HD
    • Requires a donor which can take time
    • More risky for certain groups (aged, diabetics, etc.)
    • Patient must take rejection medication for life
    • Rejection medication has side effects
    • Thrombosis
    • Hemorrhage
    • Ureteral blockage
    • Infection
    • Organ rejection
    • Death
    • Myocardial infarction
    • Stroke
    • Comprehensive Conservative Care
    • Less imposed symptom burden than dialysis
    • Preserves life satisfaction
    • May aggravate clinical condition
    • Not designed to treat, but to minimize adverse events
    • Treatment may not actually minimize risks associated with CKD

    7. Suggested Profile and Training for Users

    The catheter should be inserted, manipulated, and removed by a qualified, licensed physician or other qualified health care professional under the direction of a physician.

    8. Reference to Any Harmonized Standards and Common Specifications (CS) Applied

    Harmonized Standard or CS Revision Title or Description Level of Compliance
    EN 556-1 2001 Sterilization of medical devices. Requirements for medical devices to be designated "STERILE". Requirements for terminally sterilized medical devices Full
    EN ISO 10555-1 2013+A1:2017 Intravascular catheters. Sterile and single-use catheters. General requirements Full
    EN ISO 10555-3 2013 Intravascular catheters. Sterile and single-use catheters. Central venous catheters Full
    EN ISO 10993-1 2020 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process Full
    EN ISO 10993-7 2008+ A1:2022 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals — Amendment 1: Applicability of allowable limits for neonates and infants Full
    EN ISO 10993-18 2020 Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process Full
    EN ISO 11070 2014+A1:2018 Sterile single-use intravascular introducers, dilators and guidewires Full
    EN ISO 11135 2014 + A1: 2019 Sterilization of health-care products. Ethylene oxide. Requirements for the development, validation and routine control of a sterilization process for medical devices Full
    EN ISO 11138-1 2017 Sterilization of health care products — Biological indicators Part 1: General requirements Full
    EN ISO 11138-2 2017 Sterilization of health care products—Biological indicators—Part 2: Biological indicators for ethylene oxide sterilization processes Full
    EN ISO 11138-7 2019 Sterilization of health care products. Biological indicators - Guidance for the selection, use and interpretation of results Full
    EN ISO 11140-1 2014 Sterilization of health care products — Chemical indicators Part 1: General requirements Full
    EN ISO 11607-1 2020 Packaging for terminally sterilized medical devices. Requirements for materials, sterile barrier systems and packaging systems Full
    EN ISO 11607-2 2020 Packaging for terminally sterilized medical devices. Validation requirements for forming, sealing and assembly processes Full
    EN ISO 11737-1 2018 + A1: 2021 Sterilization of health care products. Microbiological methods. Determination of a population of microorganisms on products Full
    EN ISO 13485 2016 + A11: 2021 Medical Devices – Quality Management system – Requirements for Regulatory Purposes Full
    EN ISO 14155 2020 Clinical investigation of medical devices for human subjects — Good clinical practice Full
    EN ISO 14644-1 2015 Cleanrooms and associated controlled environments — Part 1: Classification of air cleanliness by particle concentration Full
    EN ISO 14644-2 2015 Cleanrooms and associated controlled environments — Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration Full
    EN ISO 14971 2019+A11:2021 Medical devices. Application of risk management to medical devices Full
    EN ISO 15223-1 2021 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements Full
    EN ISO/IEC 17025 2017 General requirements for the competence of testing and calibration laboratories Full
    PD CEN ISO/TR 20416 2020 Medical devices — post-market surveillance for manufacturers Full
    EN ISO 20417 2021 Medical devices - Information to be supplied by the manufacturer. Full
    EN 62366-1 2015 + A1: 2020 Medical devices — Part 1: Application of usability engineering to medical devices Full
    ISO 7000 2019 Graphical symbols for use on equipment. Registered symbols Partial
    ISO 594-1 1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 1: General requirements Full
    ISO 594-2 1998 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 2: Lock Fittings Full
    MEDDEV 2.7.1 Rev 4 Clinical Evaluation: A Guide for Manufacturers and Notified Bodies Under Directives 93/42/EEC and 90/385/EEC Full
    MEDDEV 2.12/2 Rev. 2 GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP STUDIES A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES Full
    MDCG 2020-6 2020 Clinical evidence needed for medical devices previously CE marked under Directives 93/42/EEC or 90/385/EEC Full
    MDCG 2020-7 2020 Post-market clinical follow-up (PMCF) Plan Template A guide for manufacturers and notified bodies Full
    MDCG 2020-8 2020 Post-market clinical follow-up (PMCF) Evaluation Report Template A guide for manufacturers and notified bodies Full
    MDCG 2018-1 Rev. 4 Guidance on BASIC UDI-DI and changes to UDI-DI Full
    MDCG 2019-9 2022 Summary of safety and clinical performance Full
    ASTM D 4169-22 2022 Standard Practices for Performance Testing of Shipping Containers and Systems. Full
    ASTM F2096-11 2019 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) Full
    ASTM F2503-20 2020 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment Full
    ASTM F640-20 2020 Standard Test Methods for determining Radiopacity for Medical Use Full
    ASTM D4332-14 2014 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing Full
    Regulation (EU) 2017/745 2017 Regulation (EU) 2017/745 of the European Parliament and of the Council Full

    Revision History

    Revision Date CR# Author Description of Changes Validated
    1 07NOV2022 27445 KO Initial Implementation of SSCP No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    2 06JUN2023 28181 KO Addition of planned PMCF activity PMCF_STHD_241; updated language throughout the patient section to enhance readability No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    3 15APR2024 29025 GM Update SSCP to include the addition of planned PMCF activity Truveta Data Queries and Retrospective Analysis and updated post market surveillance information No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    4 16SEP2024 29468 GM Update in accordance with CER-020 Revision D No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device

    Version 5.00 of Medical Components, Inc. Template QA-CL-200-1