Summary of Safety and Clinical Performance

Duo-Flow® Side x Side Catheter

SSCP Document Number: SSCP-029
Revision Number: 3
Revision Date: 16-Sep-24

Important Information

This Summary of Safety and Clinical Performance (SSCP) is intended to provide public access to an updated summary of the main aspects of the safety and clinical performance of the device. This SSCP is not intended to replace the Instructions for Use as the main document to ensure the safe use of the device, nor is it intended to provide diagnostic or therapeutic suggestions to intended users or patients.

Applicable Documents

Document Type Document Title / Number
Design History File (DHF) 02030, 03010, 17009, 17009-A1, 11025
‘MDR Documentation’ File Number TD-029

1. Device Identification and General Information

Device Trade Name(s): Duo-Flow® Side x Side Catheter

Manufacturer Name and Address: Medical Components, Inc. 1499 Delp Drive Harleysville, PA 19438 USA

Manufacturer Single Registration Number (SRN): US-MF-000008230

Basic UDI-DI: 00884908303MW

Medical Device Nomenclature: F900201 – Temporary Hemodialysis Catheters and Kits

Class of Device: III

Date First CE Certificate Issued: Nov-97

Authorized Representative Name and SRN: Gerhard Frömel European Regulatory Expert Medical Product Service GmbH (MPS) Borngasse 20 35619 Braunfels, Germany SRN: DE-AR-000005009

Notified Body Name and Single Identification Number: BSI Netherlands NB2797

Device Grouping and Variants

The devices in scope of this document are all short-term hemodialysis catheter sets. The device part numbers are organized into variant categories. These devices are distributed as procedure trays, in various configurations inclusive of accessories and adjunctive devices (see section “Accessories intended for use in combination with the Device”).

Variant Devices:

Variant Devices:
Variant Description Part Number(s) Explanation of Multiple Part Numbers
11F x 12cm Curved Extensions Duo-Flow Side x Side 1176
11F x 12cm Straight Duo-Flow Side x Side 1084
11F x 13.5cm Curved Extensions Duo-Flow Side x Side 10540
11F x 15cm Curved Extensions Duo-Flow Side x Side 1174
11F x 15cm Straight Duo-Flow Side x Side 1085
11F x 20cm Curved Extensions Duo-Flow Side x Side 1175
11F x 20cm Straight Duo-Flow Side x Side 1086
11F x 24cm Straight Duo-Flow Side x Side 1246
12F x 13.5cm Curved Extensions Duo-Flow Side x Side 10011-8135-105C
12F x 13cm Curved Extensions Duo-Flow Side x Side 10011-813-100C 10011-813-105C
12F x 13cm Straight Duo-Flow Side x Side 10011-813-100 10011-813-105
12F x 15cm Curved Extensions Duo-Flow Side x Side 10011-815-100C 10011-815-105C
12F x 15cm Straight Duo-Flow Side x Side 10011-815-100 10011-815-105
12F x 20cm Curved Extensions Duo-Flow Side x Side 10011-820-100C 10011-820-105C
12F x 20cm Straight Duo-Flow Side x Side 10011-820-100 10011-820-105
12F x 24cm Curved Extensions Duo-Flow Side x Side 10011-824-100C 10011-824-105C
12F x 24cm Straight Duo-Flow Side x Side 10011-824-100 10011-824-105
14F x 13cm Curved Extensions Duo-Flow Side x Side 10067-813-100C 10067-813-105C
14F x 13cm Straight Duo-Flow Side x Side 10067-813-100 10067-813-105
14F x 15cm Curved Extensions Duo-Flow Side x Side 10067-815-100C 10067-815-105C
14F x 15cm Straight Duo-Flow Side x Side 10067-815-100 10067-815-105
14F x 20cm Curved Extensions Duo-Flow Side x Side 10067-820-100C 10067-820-105C
14F x 20cm Straight Duo-Flow Side x Side 10067-820-100 10067-820-105
14F x 24cm Curved Extensions Duo-Flow Side x Side 10067-824-100C 10067-824-105C
14F x 24cm Straight Duo-Flow Side x Side 10067-824-100 10067-824-105
9F x 12cm Curved Extensions Duo-Flow Side x Side 1177
9F x 12cm Straight Duo-Flow Side x Side 1124
9F x 15cm Curved Extensions Duo-Flow Side x Side 1178
9F x 15cm Straight Duo-Flow Side x Side 1125
9F x 20cm Straight Duo-Flow Side x Side 1126
Variant Devices:
Variant Description Part Number(s) Explanation of Multiple Part Numbers

Procedure Trays:

Procedure Trays:
Catalog Code Part Number Description
ARD1213C 10011-813-105C 12F X 13Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1213S 10011-813-105 12F X 13Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD1215C 10011-815-105C 12F X 15Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1215S 10011-815-105 12F X 15Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD1220C 10011-820-105C 12F X 20Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1220S 10011-820-105 12F X 20Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD1224C 10011-824-105C 12F X 24Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1224S 10011-824-105 12F X 24Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD1235C 10011-8135-105C 12Fx13.5Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1413C 10067-813-105C 14F X 13Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1413S 10067-813-105 14F X 13Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD1415C 10067-815-105C 14F X 15Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1415S 10067-815-105 14F X 15Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD1420C 10067-820-105C 14F X 20Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1420S 10067-820-105 14F X 20Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD1424C 10067-824-105C 14F X 24Cm Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
ARD1424S 10067-824-105 14F X 24Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
ARD912S 1124 9F X 12Cm Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
DL 11/15 1085 11F X 15Cm Nikkiso Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
DL 11/20 1086 11F X 20Cm Nikkiso Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
DL 9/15 1125 9F X 15Cm Nikkiso Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
NDDLPC15 10011-815-112C 12F X 15Cm Nipro Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
NDDLPC20 10011-820-112C 12F X 20Cm Nipro Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
NDDLS20 10011-820-112 12F X 20Cm Nipro Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
NIDLC15K 10011-815-112C 12F X 15Cm Nipro Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
NIDLC20K 10011-820-112C 12F X 20Cm Nipro Jet Cath® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
NIDLS15K 10011-815-112 12F X 15Cm Nipro Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
NIDLS20K 10011-820-112 12F X 20Cm Nipro Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
NIDLS24K 10011-824-112 12F X 24Cm Nipro Jet Cath® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP114IJS= 1176 11F X 12Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP114MT= 1084 11F X 12Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP115IJS= 10540 11F X 13.5Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP116IJS= 1174 11F X 15Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP116MT= 1085 11F X 15Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP118IJS= 1175 11F X 20Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP118MT= 1086 11F X 20Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP119MT= 1246 11F X 24Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP125IJSA 10011-813-100C 12F X 13Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP125MTA 10011-813-100 12F X 13Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP126IJSA 10011-815-100C 12F X 15Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP126MTA 10011-815-100 12F X 15Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP128IJSA 10011-820-100C 12F X 20Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP128MTA 10011-820-100 12F X 20Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP129IJSA 10011-824-100C 12F X 24Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP129MTA 10011-824-100 12F X 24Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP145IJSA 10067-813-100C 14F X 13Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP145MTA 10067-813-100 14F X 13Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP146IJSA 10067-815-100C 14F X 15Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP146MTA 10067-815-100 14F X 15Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP148IJSA 10067-820-100C 14F X 20Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP148MTA 10067-820-100 14F X 20Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP149IJSA 10067-824-100C 14F X 24Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP149MTA 10067-824-100 14F X 24Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP94IJS= 1177 9F X 12Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
XTP94MT= 1124 9F X 12Cm Duo-Flow® Side X Side Double Lumen Hemodialysis Catheter Basic Set
XTP96IJS= 1178 9F X 15Cm Duo-Flow® Side X Side Double Lumen Curved Extension Hemodialysis Catheter Basic Set
Procedure Trays:
Catalog Code Part Number Description

Configurations of Procedure Trays:

Configuration Type Kit Components
Basic Set (1) Catheter (1) Guidewire (1) Guidewire Advancer (1) Needle (1) Scalpel (1) Syringe (1) Dilator (2) End Cap

2. Intended Use of the Device

Intended Purpose: The Duo-Flow® Side x Side Catheters are intended for use in adult patients with Acute Kidney Injury (AKI) or Chronic Kidney Disease (CKD) for whom immediate central venous vascular access for short-term hemodialysis is deemed necessary based on the direction of a qualified, licensed physician. The catheter is intended to be used under the regular review and assessment of qualified health professionals. This catheter is for Single Use Only.

Indication(s): The Duo-Flow® Side x Side Catheter is indicated for short-term use where vascular access is required for less than 14 days for the purpose of hemodialysis.

Target Population(s): Duo-Flow® Side x Side catheters are intended for use in adult patients with Acute Kidney Injury (AKI) or Chronic Kidney Disease (CKD) for whom immediate central venous vascular access for short-term hemodialysis is deemed necessary based on the direction of a qualified, licensed physician. The catheter is not intended for use in pediatric patients.

Contraindications and/or Limitations:

  • Known or suspected allergies to any of the components of the catheter or the kit.
  • This device is contraindicated for patients exhibiting severe, uncontrolled coagulopathy or thrombocytopenia.

3. Device Description

Device Image

Device Name: Duo-Flow® Side x Side Catheter

Description of Device: The Duo-Flow® Side x Side Catheter removes and returns blood through two segregated lumen passages. Each lumen is connected to an extension line with color-coded female luer connectors. The transition between lumen and extension is housed within a molded hub. Both arterial and venous lumens contain side-holes. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. The catheter is available with straight or curved extensions in a variety of French sizes and lengths to accommodate physician preference and clinical needs. The curved extensions devices are not suitable for femoral insertion.

Device Image

Device Name: Duo-Jet® Side x Side Catheter

Description of Device: The Duo-Jet® Side x Side Catheter removes and returns blood through two segregated lumen passages. Each lumen is connected to an extension line with color-coded female luer connectors. The transition between lumen and extension is housed within a molded hub. Both arterial and venous lumens contain side-holes. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. The catheter is available with straight or curved extensions in a variety of French sizes and lengths to accommodate physician preference and clinical needs. The curved extensions devices are not suitable for femoral insertion. Nipro Duo-Flow® Side x Side Catheter The Nipro Duo-Flow® Side x Side Catheter removes and returns blood through two segregated lumen passages. Each lumen is connected to an extension line with color-coded female luer connectors. The transition between lumen and extension is housed within a molded hub. Both arterial and venous lumens contain side-holes. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. The catheter is available with straight or curved extensions in a variety of lengths to accommodate physician preference and clin Nikkiso Duo-Flow® Side x Side Catheter The Nikkiso Duo-Flow® Side x Side Catheter removes and returns blood through two segregated lumen passages. Each lumen is connected to an extension line with color-coded female luer connectors. The transition between lumen and extension is housed within a molded hub. Both arterial and venous lumens contain side-holes. The catheter incorporates Barium Sulphate to facilitate visualization under fluoroscopy or Xray. The catheter is available in a variety of French sizes and lengths to accommodate physician preference and clinical needs.

Materials / Substances in Contact with Patient Tissue:

The percentage ranges in the table below are based on the weights of the 9F x 12cm straight catheter (7.87g) and the 14F x 24cm catheter with curved extensions (11.61g).

Material % Weight (w/w)
Polyurethane 37.37 - 48.13
Acetal co-polymer 20.64 - 24.95
Polyvinyl chloride 16.80 - 24.83
Acrylonitrile Butadiene Styrene 6.33 - 9.34
Vythene 0 - 5.76
Barium sulfate 1.16 - 5.23

Material % Weight (w/w)

Information on Medicinal Substances in the Device:

How the Device Achieves its Intended Mode of Action: Hemodialysis catheters are centrally placed access tubes. A typical hemodialysis catheter uses a thin, flexible tube. The tube has two openings. The tube goes into a large vein. The vein is usually the internal jugular vein. Blood withdraws through one lumen of the catheter. The blood flows to the dialysis machine through a separate tubing set. The blood is then processed and filtered. The blood returns to the patient through the second lumen. This device is used when dialysis must start at once. Patients may not have a functioning AV fistula or graft. Catheter hemodialysis normally happens on a short-term basis.

Sterilization Information: Contents sterile and non-pyrogenic in unopened, undamaged package. Sterilized by Ethylene Oxide.

Previous Generations / Variants:

Name of Previous Generation Differences from Current Device
N/A N/A

Accessories Intended for Use in Combination with the Device:

Name of Accessory Description of Accessory
Guidewire For general intravascular use to facilitate the selective placement of medical devices in the vessel anatomy.
Guidewire Advancer Aid for introduction of guidewire into target vein.
Introducer Needle Used for the percutaneous introduction of guidewires.
Scalpel A cutting device during surgical, pathology and minor medical procedures
Dilator Designed for percutaneous entry into a vessel in order to enlarge the opening of the vessel for the placement of a catheter in a vein.
End Cap To keep clean and protect catheter luer between treatments.

Other Devices or Products Intended for Use in Combination with the Device:

Name of Device or Product Description of Device or Product
Syringe Attached to introducer needle to help capture blood return once introducer needle perforates targeted vein, prevent air embolism

4. Risks and Warnings

Residual Risks and Undesirable Effects: All surgical procedures carry risk. Medcomp® has implemented risk management processes to proactively find and mitigate these risks as far as possible without adversely affecting the benefit-risk profile of the device. After mitigation, residual risks and the possibility of adverse events from use of this product remain. Medcomp® has determined that all residual risks are acceptable when considered with respect to the expected clinical benefits of the Duo-Flow® Side x Side Catheter and the benefits of other similar hemodialysis devices.

Residual Harm Type Possible Adverse Events Associated with Harm
Allergic Reaction Allergic Reaction Intolerance Reaction to Implanted Device
Bleeding Bleeding (May be severe) Exsanguination Femoral Artery Bleed Hematoma Hemorrhage Retroperitoneal Bleed
Cardiac Event Cardiac Arrhythmia Cardiac Tamponade
Embolism Air Embolus
Infection Bacteremia Endocarditis Exit Site Infection Septicemia
Perforation Inferior Vena Cava Puncture Laceration of the Vessel Perforation of the Vessel Pneumothorax Right Atrial Puncture Subclavian Artery Puncture Superior Vena Cava Puncture
Stenosis Venous Stenosis
Tissue Injury Brachial Plexus Injury Exit Site Necrosis Mediastinal Injury Pleural Injury
Thrombosis Central Venous Thrombosis Lumen Thrombosis Subclavian Vein Thrombosis Vascular Thrombosis
Miscellaneous Complications Catheter Dysfunction Femoral Nerve Damage Hemothorax Malposition Thoracic Duct Laceration
Quantification of Residual Risks
PMS Complaints 01 January 2019 – 31 December 2023 PMCF Events
Units Sold: 1,204,381 Units Studied: 94
Patient Residual Harm Category % of Devices % of Devices
Allergic Reaction Not Reported Not Reported
Bleeding 0.00008% Not Reported
Cardiac Event Not Reported Not Reported
Embolism Not Reported Not Reported
Infection 0.00033% 3.19%
Perforation 0.00008% Not Reported
Stenosis Not Reported Not Reported
Tissue Injury Not Reported Not Reported
Thrombosis Not Reported 1.06%

Warnings and Precautions:

Warnings listed for the Duo-Flow® Side x Side Catheter are as follows:

  • Warnings listed for the Duo-Flow® Side x Side Catheter are as follows:
  • Do not insert catheter in thrombosed vessels.
  • Do not advance the guidewire or catheter if unusual resistance is encountered.
  • Do not insert or withdraw the guidewire forcibly from any component. If the guidewire becomes damaged, guidewire and any associated componentry must be removed together.
  • Do not resterilize the catheter or accessories by any method.
  • Contents sterile and non-pyrogenic in unopened, undamaged package. STERILIZED BY ETHYLENE OXIDE
  • Do not re-use catheter or accessories as there may be a failure to adequately clean and decontaminate the device which may lead to contamination, catheter degradation, device fatigue, or endotoxin reaction.
  • Do not use catheter or accessories if package is opened or damaged.
  • Do not use catheter or accessories if any sign of product damage is visible or the use-by date has passed.
  • Do not use sharp instruments near the extension tubing or catheter lumen.
  • Do not use scissors to remove dressing. Precautions listed for Duo-Flow® Side x Side Catheter are as follows:
  • Examine catheter lumen and extensions before and after each treatment for damage.
  • To prevent accidents, ensure the security of all caps and bloodline connections prior to and between treatments.
  • Use only Luer Lock (threaded) Connectors with this catheter.
  • In the rare event that a hub or connector separates from any component during insertion or use, take all necessary steps and precautions to prevent blood loss or air embolism and remove the catheter.
  • Before attempting catheter insertion, ensure that you are familiar with the potential complications and their emergency treatment should any of them occur.
  • Repeated overtightening of bloodlines, syringes, and caps will reduce connector life and could lead to potential connector failure.
  • The catheter will be damaged if clamps other than what is provided with this kit are used.
  • Avoid clamping near the Luer Lock and hub of the catheter. Clamping of the tubing repeatedly in the same location may weaken tubing.

Other Relevant Aspects of Safety: For a period of 01 January 2019 to 31 December 2023 there were 139 complaints for 1,204,381 units sold, giving an overall complaint rate of 0.012%. There were no death-related events during the review period. No events resulted in recalls during the review period.

5. Summary of Clinical Evaluation and Post-Market Clinical Follow-Up (PMCF)

Summary of Clinical Data Related to Subject Device

Specific Case Numbers (Mixed Cohort Case Numbers) Identified and Used for Clinical Performance Evaluation
Product Family Clinical Literature PMCF Data Total Cases User Survey Responses
0 (& 67 Mixed Cohort Cases) 94 94 (& 67 Mixed Cohort Cases) 1

Clinical performance was measured using parameters including but not limited to dwell time, catheter insertion outcomes, and adverse event rates. Critical clinical parameters extracted from these studies met standards set forth in the guidelines for the State of the Art. There were no unforeseen adverse events or other high occurrences of adverse events detected in any of the clinical activities. Medcomp® STHD catheters are subjected to, and must pass, simulated use testing intended to replicate 30 days use as part of device development. The Duo-Flow® Side x Side Catheter passed this testing. Clinical guidelines recommend to limit the use of temporary, noncuffed, nontunneled dialysis catheters to a maximum of 2 weeks (KDOQI 2019), however, duration of use of these catheters has varied in available clinical evidence identified by the manufacturer to date. Although Medcomp® catheters materials contain non-degradable polymers, fully functional catheters may be removed for other reasons, such as intractable infection or change of therapy. Published clinical literature does not always focus on the physical lifetime of a catheter for these reasons. In the case of the Duo-Flow® Side x Side Catheter, 14 cases had a mean dwell time of 33.07 days [95% CI: 25.50 – 40.64 days] duration of use that has been found in clinical use reported to date. Based on this information, the Duo-Flow® Side x Side catheter has a 30 day lifetime; however, the decision to remove and/or replace the catheter should be based on clinical performance and need, and not any predetermined point in time.

Summary of Clinical Data Related to the Equivalent Device

Clinical evidence from published literature and PMCF activities has been generated specific to known and unknown variants of the subject device. An equivalency rationale within the manufacturer’s technical documentation demonstrates that the clinical evidence available for these variants is representative of the range of device variants in the device family. There are no clinical or biological differences between variants within the subject device family, and the potential impact of the technical differences has been rationalized.

Summary of Clinical Data from Pre-Market Investigations (if applicable)

No pre-market clinical devices were used for the device’s clinical evaluation.

Summary of Clinical Data from Other Sources

Source:Summary of Published Literature

Previous clinical evidence literature searches found one published literature article representing 67 mixed cohort cases inclusive of the Duo-Flow® Side x Side device family. The most recent clinical evidence search found no published literature articles relating to the Duo-Flow® Side x Side device family. The article included a retrospective study (Silva et al.). Bibliography: de Jesus-Silva SG, Oliveira JDS, Ramos KTF et al. Analysis of infection rates and duration of short and long-term hemodialysis catheters in a teaching hospital. Jornal vascular brasileiro. 2020;19:e20190142.

• Source:PMCF_Medcomp_211

The Medcomp User Survey acquired responses from healthcare personnel familiar with any number of Medcomp’s product offerings. 20 respondents responded that they or their facility have used Medcomp short-term hemodialysis catheters, with 1 of those respondents using the Duo-Flow® Side x Side device family. There were no differences in mean user sentiments within short-term hemodialysis catheters across State of the Art Performance and Safety Outcome Measures or between device types relating to safety or performance. The following data points were collected from users of Medcomp short-term hemodialysis catheters (n=20):

  • (Mean Likert Scale Response) Catheters function as intended – 4.8 / 5
  • (Mean Likert Scale Response) Packaging allows for aseptic presentation – 4.9 / 5
  • (Mean Likert Scale Response) Benefit outweighs the risk – 4.7 / 5
  • Dwell Time (n=19) – 15.74 days (95%CI: 6.3 – 25.1) The following data points were collected from users of Medcomp Duo-Flow Side x Side catheters (n=1):
  • (Mean Likert Scale Response) Catheters function as intended – 5 / 5
  • (Mean Likert Scale Response) Packaging allows for aseptic presentation – 5 / 5
  • (Mean Likert Scale Response) Benefit outweighs the risk – 5 / 5
  • Dwell Time (n=1) – 21 days
  • • Source:PMCF_STHD_211 (Retrospective Patient-Level Usage Data Survey)

    The Short-Term Hemodialysis (STHD) Product Line Data Collection Survey aimed to assess safety and performance outcome information for all variants of Medcomp STHD catheters. 19 survey responses were collected from 10 countries representing 381 device cases. 85 Duo-Flow® Side x Side cases inclusive of several variant devices across French size (11F, 12F), length (15cm and 20cm), and catheter geometry (straight and curved extensions) were collected. The following outcome measures were confirmed to be within State of the Art safety and performance outcome measures from published literature for Medcomp Duo-Flow® Side x Side devices:

  • Dwell Time – 33.07 Days (95%CI: 25.50 – 40.64)
  • Procedural Outcomes – 100%
  • Catheter Related Blood Stream Infection – 4.31 per 1,000 catheter days (95%CI: 0 – 10.31)
  • Catheter Associated Venous Thrombus – 4.31 per 1,000 catheter days (95%CI: 0 – 10.31)
  • Exit Site Infection – 4.31 per 1,000 catheter days (95%CI: 0 – 10.31)
  • • Source:PMCF_Infusion_211 (Retrospective Patient-Level Usage Data Survey)

    The Infusion Product Line Data Collection Survey aimed to assess safety and performance outcome information for all variants of Medcomp Infusion Ports, PICCs, Midlines, and CVCs. 70 survey responses were collected from 17 countries representing 471 device cases. 7 Duo-Flow® Side x Side cases, all described as 12F, inclusive of 15cm and 20cm variants were collected. The following outcome measures were collected for Medcomp Duo-Flow® Side x Side devices:

  • Dwell Time – 44.5 days (Range: 18 – 71 days)
  • Procedural Outcomes – 100%
  • Catheter Related Blood Stream Infection – No Events Reported
  • Catheter Associated Venous Thrombus – No Events Reported
  • Exit Site Infection – No Events Reported
  • • Source:PMCF_STHD_242 (Short-Term Hemodialysis Truveta Data Analysis)

    The Short-Term Hemodialysis (STHD) Truveta data analysis assessed safety and performance outcome information for Medcomp® and competitor devices present in Truveta Studio. Truveta data comes from a growing collective of more than 30 health systems that provide 17% of the daily clinical care across all 50 U.S. states from 800 hospitals and 20,000 clinics, representing the full diversity of the United States. The population used for data analysis was derived utilizing Truveta Studio’s proprietary coding language (Prose) and unique device identifier (UDI) codes representing all saleable Medcomp® STHD devices and STHD devices distributed and/or manufactured by other companies. 2 Duo-Flow® Side x Side cases inclusive of several variant devices were collected. Cases were described as 11F and Curved Extension and Straight Cases included multiple French sizes (11F, 12F), configurations (straight, curved extension), and lengths (15cm), representation of 15cm length catheters. The following State of the Art safety and performance outcome measures were observed for Medcomp Duo-Flow® Side x Side devices:

  • Catheter Related Blood Stream Infection – 35.71 per 1,000 catheter days (95%CI: 0.9 – 198.99)
  • Catheter Associated Venous Thrombus – 0 per 1,000 catheter days (95%CI: 0 – 131.75)
  • Exit Site Infection – 0 per 1,000 catheter days (95%CI: 0 – 131.75) The catheter brand logistic regression model did not find that any Medcomp® catheter brands were statistically significantly associated with the incidence of CRBSI. The brand agnostic logistic regression found that Triple Lumen catheters OR: 1.63 (95%CI: 1.17 – 2.28) (as compared to the reference category of Double Lumen catheters) and Pre-Curved catheters OR: 7.26 (95%CI: 1.32 – 32.69) (as compared to the reference category of straight catheters) were statistically significantly associated with the incidence of CRBSI.
  • Overall Summary of Clinical Safety and Performance

    Outcome Benefit/Risk Acceptability Criteria Desired Trend Clinical Literature (Subject Device) PMCF Data (Subject Device)
    Performance Outcomes
    Dwell Time Greater than 8 days + ND*
    33.07 days (95%CI
    Procedural Outcomes Greater than 95% + ND*
    100% (PMCF_STHD_211 & PMCF_Infusion_211 ) Likert Scale Response 5 / 5 (PMCF_Medcomp_211)**
    Safety Outcomes
    Catheter Related Blood Stream Infection (CRBSI) Less than 7.8 incidents of CRBSI per 1,000 catheter days - ND*
    4.31 per 1,000 catheter days (95%CI
    Exit Site Infection Rate Less than 3.5 incidents of exit site infection per 1,000 catheter days - ND*
    4.31 per 1,000 catheter days (95%CI
    Catheter Associated Venous Thrombus (CAVT) Less than 11.4 incidents of CAVT per 1,000 catheter days - ND*
    4.31 per 1,000 catheter days (95%CI
    *ND = No Data on Parameter: **PMCF_Medcomp_211 asked respondents, if they agreed on a scale of 1 -5, that their experience in relation to each outcome was the same or better than the benefit/risk acceptability criteria.:
    Outcome Benefit/Risk Acceptability Criteria Desired Trend Clinical Literature (Subject Device) PMCF Data (Subject Device)
    Performance Outcomes
    Multicenter Patient-Level Case Series Collect additional clinical data on the device PMCF_STHD_241 Q4 2025
    State of the Art Literature Search Identify risks and trends with use of dialysis catheters SAP-HD Q1 2025
    Clinical Evidence Literature Search Identify risks and trends with use of the device LRP-STHD Q3 2025
    Global Trial Database Search Identify ongoing clinical trials involving Duo-Flow® Side x Side catheters N/A Q3 2025
    Safety Outcomes

    On-going or Planned Post-Market Clinical Follow-Up (PMCF)

    Activity Description Reference Timeline

    No emerging risks, complications or unexpected device failures have been detected from PMCF activities.

    6. Possible Therapeutic Alternatives

    The Kidney Disease Outcomes Quality Initiative (KDOQI) 2019 clinical practice guidelines have been used to support the below recommendations for treatments.

    Therapy Benefits Disadvantages Key Risks
    • AV Fistula
    • Permanent vascular access solution
    • Lower complication rate than hemodialysis via catheter.
    • Requires time to mature
    • Patients must sometimes self-cannulate
    • Stenosis
    • Thrombosis
    • Aneurysm
    • Pulmonary hypertension
    • Steal Syndrome
    • Septicemia
    • Hemodialysis Catheter
    • Useful for quick vascular access without AV Fistula in place
    • Can be used as a bridge dialysis method between other therapies
    • Not a permanent solution
    • Catheter dysfunction can disrupt regular treatment
    • Benefit is not equal for all patient populations
    • Post-procedural bleeding
    • Infection
    • Thrombosis
    • Decreased blood flow in dysfunctional catheter
    • Cardiovascular events
    • Fibrin sheath formation around catheter
    • Septicemia
    • Peritoneal Dialysis
    • Less restrictive diet than hemodialysis
    • Does not require hospitalization, can be done in any clean place
    • Clearance of impurities is limited by dialysate flow and peritoneal area
    • Peritonitis
    • Septicemia
    • Fluid overload
    • Kidney Transplant
    • Better quality of life compared to HD
    • Lower risk of death compared to HD
    • Fewer dietary restrictions compared to HD
    • Requires a donor which can take time
    • More risky for certain groups (aged, diabetics, etc.)
    • Patient must take rejection medication for life
    • Rejection medication has side effects
    • Thrombosis
    • Hemorrhage
    • Ureteral blockage
    • Infection
    • Organ rejection
    • Death
    • Myocardial infarction
    • Stroke
    • Comprehensive Conservative Care
    • Less imposed symptom burden than dialysis
    • Preserves life satisfaction
    • May aggravate clinical condition
    • Not designed to treat, but to minimize adverse events
    • Treatment may not actually minimize risks associated with CKD

    7. Suggested Profile and Training for Users

    The catheter should be inserted, manipulated, and removed by a qualified, licensed physician or other qualified health care professional under the direction of a physician.

    8. Reference to Any Harmonized Standards and Common Specifications (CS) Applied

    Harmonized Standard or CS Revision Title or Description Level of Compliance
    EN 556-1 2001 Sterilization of medical devices. Requirements for medical devices to be designated "STERILE". Requirements for terminally sterilized medical devices Full
    EN ISO 10555-1 2013+A1:2017 Intravascular catheters. Sterile and single use catheters. General requirements Full
    EN ISO 10555-3 2013 Intravascular catheters. Sterile and single-use catheters. Central venous catheters Full
    EN ISO 10993-1 2020 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process Full
    EN ISO 10993-7 2008+ A1:2022 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals — Amendment 1: Applicability of allowable limits for neonates and infants Full
    EN ISO 10993-18 2020 Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process Full
    EN ISO 11070 2014+A1:2018 Sterile single-use intravascular introducers, dilators and guidewires Full
    EN ISO 11135 2014 + A1: 2019 Sterilization of health-care products. Ethylene oxide. Requirements for the development, validation and routine control of a sterilization process for medical devices Full
    EN ISO 11138-1 2017 Sterilization of health care products — Biological indicators Part 1: General requirements Full
    EN ISO 11138-2 2017 Sterilization of health care products—Biological indicators—Part 2: Biological indicators for ethylene oxide sterilization processes Full
    EN ISO 11138-7 2019 Sterilization of health care products. Biological indicators - Guidance for the selection, use and interpretation of results Full
    EN ISO 11140-1 2014 Sterilization of health care products — Chemical indicators Part 1: General requirements Full
    EN ISO 11607-1 2020 Packaging for terminally sterilized medical devices. Requirements for materials, sterile barrier systems and packaging systems Full
    EN ISO 11607-2 2020 Packaging for terminally sterilized medical devices. Validation requirements for forming, sealing and assembly processes Full
    EN ISO 11737-1 2018 + A1: 2021 Sterilization of health care products. Microbiological methods. Determination of a population of microorganisms on products Full
    EN ISO 13485 2016 + A11: 2021 Medical Devices – Quality Management system – Requirements for Regulatory Purposes Full
    EN ISO 14155 2020 Clinical investigation of medical devices for human subjects — Good clinical practice Full
    EN ISO 14644-1 2015 Cleanrooms and associated controlled environments — Part 1: Classification of air cleanliness by particle concentration Full
    EN ISO 14644-2 2015 Cleanrooms and associated controlled environments — Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration Full
    EN ISO 14971 2019+A11:2021 Medical devices. Application of risk management to medical devices Full
    EN ISO 15223-1 2021 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements Full
    EN ISO/IEC 17025 2017 General requirements for the competence of testing and calibration laboratories Full
    PD CEN ISO/TR 20416 2020 Medical devices — post-market surveillance for manufacturers Full
    EN ISO 20417 2021 Medical devices - Information to be supplied by the manufacturer. Full
    EN 62366-1 2015 + A1: 2020 Medical devices — Part 1: Application of usability engineering to medical devices Full
    ISO 7000 2019 Graphical symbols for use on equipment. Registered symbols Partial
    ISO 594-1 1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 1: General requirements Full
    ISO 594-2 1998 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 2: Lock Fittings Full
    MEDDEV 2.7.1 Rev 4 Clinical Evaluation: A Guide for Manufacturers and Notified Bodies Under Directives 93/42/EEC and 90/385/EEC Full
    MEDDEV 2.12/2 Rev. 2 GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP STUDIES A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES Full
    MDCG 2020-6 2020 Clinical evidence needed for medical devices previously CE marked under Directives 93/42/EEC or 90/385/EEC Full
    MDCG 2020-7 2020 Post-market clinical follow-up (PMCF) Plan Template A guide for manufacturers and notified bodies Full
    MDCG 2020-8 2020 Post-market clinical follow-up (PMCF) Evaluation Report Template A guide for manufacturers and notified bodies Full
    MDCG 2018-1 Rev. 4 Guidance on BASIC UDI-DI and changes to UDI-DI Full
    MDCG 2019-9 2022 Summary of safety and clinical performance Full
    ASTM D 4169-22 2022 Standard Practices for Performance Testing of Shipping Containers and Systems. Full
    ASTM F2096-11 2019 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) Full
    ASTM F2503-20 2020 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment Full
    ASTM F640-20 2020 Standard Test Methods for determining Radiopacity for Medical Use Full
    ASTM D4332-14 2014 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing Full
    Regulation (EU) 2017/745 2017 Regulation (EU) 2017/745 of the European Parliament and of the Council Full

    Revision History

    Revision Date CR# Author Description of Changes Validated
    1 07NOV2022 27445 KO Initial Implementation of SSCP No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    2 19SEP2023 28471 GM Updated SSCP in accordance with CER-029_C which includes the addition of planned PMCF activities PMCF_STHD_241 and Truveta Data Analysis; updated language throughout the patient section to enhance readability No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    3 16SEP2024 29465 GM Update in accordance with CER-029 Revision D No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device

    Version 5.00 of Medical Components, Inc. Template QA-CL-200-1