Summary of Safety and Clinical Performance

T-3® CT Catheter

SSCP Document Number: SSCP-026
Revision Number: 4
Revision Date: 15APR2024

Important Information

This Summary of Safety and Clinical Performance (SSCP) is intended to provide public access to an updated summary of the main aspects of the safety and clinical performance of the device. This SSCP is not intended to replace the Instructions for Use as the main document to ensure the safe use of the device, nor is it intended to provide diagnostic or therapeutic suggestions to intended users or patients.

Applicable Documents

Document Type Document Title / Number
Design History File (DHF) 09020
‘MDR Documentation’ File Number TD-026

1. Device Identification and General Information

Device Trade Name(s): T-3® CT Catheter

Manufacturer Name and Address: Medical Components, Inc. (dba Medcomp®) 1499 Delp Drive Harleysville, PA 19438 USA

Manufacturer Single Registration Number (SRN): US-MF-000008230

Basic UDI-DI: 00884908300MQ

Medical Device Nomenclature: F900201 – Temporary Hemodialysis Catheters and Kits

Class of Device: III

Date First CE Certificate Issued: 11-Aug

Authorized Representative Name and SRN: Gerhard Frömel European Regulatory Expert Medical Product Service GmbH (MPS) Borngasse 20 35619 Braunfels, Germany SRN: US-MF-000008230

Notified Body Name and Single Identification Number: BSI Netherlands NB2797

Device Grouping and Variants

The devices in scope of this document are all short-term hemodialysis catheter sets. The device part numbers are organized into variant categories. These devices are distributed as procedure trays, in various configurations inclusive of accessories and adjunctive devices (see section “Accessories intended for use in combination with the Device”).

Variant Devices:

Variant Devices:
Variant Description Part Number(s) Explanation of Multiple Part Numbers
15.5F x 15cm Straight T-3 CT 10468-815-001
15.5F x 20cm Straight T-3 CT 10468-820-001
15.5F x 24cm Straight T-3 CT 10468-824-001
15.5F x 28cm Straight T-3 CT 10468-828-001
15.5F x 32cm Straight T-3 CT 10468-832-001
Variant Devices:
Variant Description Part Number(s) Explanation of Multiple Part Numbers
15.5F x 15cm Straight T-3 CT 10468-815-001
15.5F x 20cm Straight T-3 CT 10468-820-001
15.5F x 24cm Straight T-3 CT 10468-824-001
15.5F x 28cm Straight T-3 CT 10468-828-001
15.5F x 32cm Straight T-3 CT 10468-832-001

Procedure Trays:

Procedure Trays:
Catalog Code Part Number Description
MC013501 10468-815-001 15.5F X 15cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013502 10468-820-001 15.5F X 20cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013503 10468-824-001 15.5F X 24cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013504 10468-828-001 15.5F X 28cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013505 10468-832-001 15.5F X 32cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
Procedure Trays:
Catalog Code Part Number Description
MC013501 10468-815-001 15.5F X 15cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013502 10468-820-001 15.5F X 20cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013503 10468-824-001 15.5F X 24cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013504 10468-828-001 15.5F X 28cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set
MC013505 10468-832-001 15.5F X 32cm T-3 CT Power Injectable Triple Lumen Hemodialysis Catheter Basic Set

Configurations of Procedure Trays:

Configuration Type Kit Components
T-3® CT Basic Set (1) Catheter (1) Guidewire (1) Guidewire Advancer (1) Needle (1) Scalpel (1) Suture (1) Peelable Introducer (2) Dilator (3) End Cap

2. Intended Use of the Device

Intended Purpose: T-3® CT Catheters are intended for use in adult patients with Acute Kidney Injury (AKI) or Chronic Kidney Disease (CKD) for whom immediate central venous vascular access for short-term hemodialysis, intravenous administration of fluids or medications, and power injection of contrast media is deemed necessary based on the direction of a qualified, licensed physician. The catheter is intended to be used under the regular review and assessment of qualified health professionals. This catheter is for Single Use Only.

Indication(s): The T-3® CT Catheter is indicated for short-term use where vascular access is required for less than 14 days for the purpose of hemodialysis. The third internal lumen is indicated for intravenous administration of fluids or medications and power injection of contrast media.

Target Population(s): T-3® CT Catheters are intended for use in adult patients with Acute Kidney Injury (AKI) or Chronic Kidney Disease (CKD) for whom immediate central venous vascular access for short-term hemodialysis, intravenous administration of fluids or medications, and power injection of contrast media is deemed necessary based on the direction of a qualified, licensed physician. The catheter is not intended for use in pediatric patients.

Contraindications and/or Limitations:

  • Known or suspected allergies to any of the components of the catheter or the kit.
  • This device is contraindicated for patients exhibiting severe, uncontrolled coagulopathy or thrombocytopenia.

3. Device Description

Device Image

Device Name: T-3® CT Catheter

Description of Device: T-3® CT Catheter:The T-3® Catheter is a short-term dialysis catheter made of thermosensitive polyurethane. The catheter has three separate lumens allowing continuous blood flow. The venous (blue) and arterial (red) lumens may be used for hemodialysis treatments. The middle (purple) lumen is independent from the two dialysis lumens, and may be used for administration of fluids or medications and power injection of contrast media. The catheter is available in a variety of lengths to accommodate physician preference and clinical needs.

Device Image

Device Name: T-3® CT Catheter

Description of Device: T-3® CT Catheter:The T-3® Catheter is a short-term dialysis catheter made of thermosensitive polyurethane. The catheter has three separate lumens allowing continuous blood flow. The venous (blue) and arterial (red) lumens may be used for hemodialysis treatments. The middle (purple) lumen is independent from the two dialysis lumens, and may be used for administration of fluids or medications and power injection of contrast media. The catheter is available in a variety of lengths to accommodate physician preference and clinical needs.

Materials / Substances in Contact with Patient Tissue:

The percentage ranges in the table below are based on the weights of the 15cm catheter (15.15g) and the 32cm catheter (17.91g).

T-3® CT Catheter
Material % Weight (w/w)
Polyurethane 64.09 - 66.55
Acetal co-polymer 16.46 - 19.46
Acrylonitrile Butadiene Styrene 5.57 - 6.59
Silicone 5.57 - 6.59
Barium sulfate 3.28 - 5.86

The percentage ranges in the table below are based on the weights of the 15cm catheter (15.15g) and the 32cm catheter (17.91g).

T-3® CT Catheter
Material % Weight (w/w)
Polyurethane 64.09 - 66.55
Acetal co-polymer 16.46 - 19.46
Acrylonitrile Butadiene Styrene 5.57 - 6.59
Silicone 5.57 - 6.59
Barium sulfate 3.28 - 5.86

Note:Per the instructions for use, the device is contraindicated for patients with known or suspected allergies to the above materials.

Information on Medicinal Substances in the Device:

How the Device Achieves its Intended Mode of Action: Hemodialysis catheters are centrally placed access tubes. A typical hemodialysis catheter uses a thin, flexible tube. The tube has two openings. The tube goes into a large vein. The vein is usually the internal jugular vein. Blood withdraws through one lumen of the catheter. The blood flows to the dialysis machine through a separate tubing set. The blood is then processed and filtered. The blood returns to the patient through the second lumen. This device is used when dialysis must start at once. Patients may not have a functioning AV fistula or graft. Catheter hemodialysis normally happens on a short-term basis.

Sterilization Information: Contents sterile and non-pyrogenic in unopened, undamaged package. Sterilized by Ethylene Oxide.

Previous Generations / Variants:

Name of Previous Generation Differences from Current Device
N/A N/A

Accessories Intended for Use in Combination with the Device:

Name of Accessory Description of Accessory
Guidewire For general intravascular use to facilitate the selective placement of medical devices in the vessel anatomy.
Guidewire Advancer Aid for introduction of guidewire into target vein.
Introducer Needle Used for the percutaneous introduction of guidewires.
Dilator Designed for percutaneous entry into a vessel in order to enlarge the opening of the vessel for the placement of a catheter in a vein.
Valved Peelable Introducer Intended to obtain central venous access to facilitate catheter insertion into the central venous system.
Scalpel A cutting device during surgical, pathology and minor medical procedures
Suture Used to secure the catheter to the body.
End Cap To keep clean and protect catheter luer between treatments.
Syringe Attached to introducer needle to help capture blood return once introducer needle perforates targeted vein, prevent air embolism
Adhesive Wound Dressing Intended to protect the catheter from contamination when not in use

Other Devices or Products Intended for Use in Combination with the Device:

Name of Device or Product Description of Device or Product
N/A N/A

4. Risks and Warnings

Residual Risks and Undesirable Effects: All surgical procedures carry risk. Medcomp® has implemented risk management processes to proactively find and mitigate these risks as far as possible without adversely affecting the benefit-risk profile of the device. After mitigation, residual risks and the possibility of adverse events from use of this product remain. Medcomp® has determined that all residual risks are acceptable when considered with respect to the expected clinical benefits of the T-3® CT Catheter and the benefits of other similar hemodialysis devices.

Residual Harm Type Possible Adverse Events Associated with Harm
Allergic Reaction Allergic Reaction
Bleeding Bleeding (May be severe)
Cardiac Event Cardiac Arrhythmia
Embolism Air Embolus
Infection Bacteremia
Perforation Inferior Vena Cava Puncture
Stenosis Venous Stenosis
Tissue Injury Brachial Plexus Injury
Thrombosis Central Venous Thrombosis
Miscellaneous Complications Catheter Dysfunction
Quantification of Residual Risks
PMS Complaints 01 January 2018 – 31 December 2023 PMCF Events
Units Sold: 21,463 Units Studied: 81
Patient Residual Harm Category % of Devices % of Devices
Allergic Reaction Not Reported Not Reported
Bleeding Not Reported Not Reported
Cardiac Event Not Reported Not Reported
Embolism Not Reported Not Reported
Infection Not Reported 1.23%
Perforation Not Reported Not Reported
Stenosis Not Reported Not Reported
Tissue Injury Not Reported Not Reported
Thrombosis Not Reported 2.47%

Warnings and Precautions:

Warnings listed for the T-3® CT Catheter are as follows:

  • Do not insert catheter in thrombosed vessels.
  • Do not advance the guidewire or catheter if unusual resistance is encountered.
  • Do not insert or withdraw the guidewire forcibly from any component. If the guidewire becomes damaged, guidewire and any associated componentry must be removed together.
  • Do not resterilize the catheter or accessories by any method.
  • Contents sterile and non-pyrogenic in unopened, undamaged package. STERILIZED BY ETHYLENE OXIDE
  • Do not re-use catheter or accessories as there may be a failure to adequately clean and decontaminate the device which may lead to contamination, catheter degradation, device fatigue, or endotoxin reaction.
  • Do not use catheter or accessories if package is opened or damaged.
  • Do not use catheter or accessories if any sign of product damage is visible or the use-by date has passed.
  • Do not use sharp instruments near the extension tubing or catheter lumen.
  • Do not use scissors to remove dressing. Precautions listed for T-3® CT Catheter are as follows:
  • Examine catheter lumen and extensions before and after each treatment for damage.
  • To prevent accidents, ensure the security of all caps and bloodline connections prior to and
  • between treatments.
  • Use only Luer Lock (threaded) Connectors with this catheter.
  • In the rare event that a hub or connector separates from any component during insertion or use, take all necessary steps and precautions to prevent blood loss or air embolism and remove the catheter.
  • Before attempting catheter insertion, ensure that you are familiar with the potential complications and their emergency treatment should any of them occur.
  • Repeated overtightening of bloodlines, syringes, and caps will reduce connector life and could lead to potential connector failure.
  • The catheter will be damaged if clamps other than what is provided with this kit are used.
  • Avoid clamping near the Luer Lock and hub of the catheter. Clamping of the tubing repeatedly in the same location may weaken tubing.
  • The third lumen of the catheter allows for intravenous administration of fluids or medications and power injection of contrast media. Refer to standards of practice and institutional policies for compatible infusion agents for central venous access.
  • Follow all contraindications, warnings, precautions, and instructions for all infusates including contrast media as specified by their manufacturer.
  • The red arterial and blue venous lumens should not be used for infusion of any infusates or contrast media as patient injury may occur.
  • The center (distal) infusion lumen should not be used for hemodialysis as insufficient treatment may occur.

Other Relevant Aspects of Safety: For a period of 01 January 2019 to 31 December 2023 there was 1 complaint for 9,842 units sold, giving an overall complaint rate of 0.01%. There were no death-related events. No events resulted in recalls during the review period.

5. Summary of Clinical Evaluation and Post-Market Clinical Follow-Up (PMCF)

Summary of Clinical Data Related to Subject Device

Specific Case Numbers (Mixed Cohort Case Numbers) Identified and Used for Clinical Performance Evaluation
Product Family Clinical Literature PMCF Data Total Cases User Survey Responses
Clinical Literature 1 (& 10 Mixed Cohort Cases)
PMCF Data 81
Total Cases 82 (& 10 Mixed Cohort Cases)
User Survey Responses 19

Clinical performance was measured using parameters including but not limited to dwell time, catheter insertion outcomes, and adverse event rates. Critical clinical parameters extracted from these studies met standards set forth in the guidelines for the State of the Art. There were no unforeseen adverse events or other high occurrences of adverse events detected in any of the clinical activities. Medcomp® STHD catheters are subjected to, and must pass, simulated use testing intended to replicate 30 days use as part of device development. The T-3® CT Catheter passed this testing. Clinical guidelines recommend to limit the use of temporary, noncuffed, nontunneled dialysis catheters to a maximum of 2 weeks (KDOQI 2019), however, duration of use of these catheters has varied in available clinical evidence identified by the manufacturer to date. Although Medcomp® catheters materials contain non-degradable polymers, fully functional catheters may be removed for other reasons, such as intractable infection or change of therapy. Published clinical literature does not always focus on the physical lifetime of a catheter for these reasons. In the case of the T-3® Catheter, 77 catheters had a mean dwell time of 11.3 days [95%CI: 9.04 – 13.61 days] duration of use that has been found in clinical use reported to date. Based on this information The T-3® CT Catheter has a 30 day lifetime; however, the decision to remove and/or replace the catheter should be based on clinical performance and need, and not any predetermined point in time.

Summary of Clinical Data Related to the Equivalent Device

Clinical evidence from published literature and PMCF activities has been generated specific to known and unknown variants of the subject device. The equivalency rationale in the updated clinical evaluation report will demonstrate that the clinical evidence available for these variants is representative of the range of device variants in the device family. There are no clinical or biological differences between variants within the subject device family, and the potential impact of the technical differences will be rationalized in the updated clinical evaluation report.

Summary of Clinical Data from Pre-Market Investigations (if applicable)

No pre-market clinical devices were used for the device’s clinical evaluation.

Summary of Clinical Data from Other Sources

Source:Summary of Published Literature

Previous clinical evidence literature searches found two published literature articles representing 1 T-3® CT device family specific case, and an additional 10 mixed cohort cases inclusive of the T-3® CT device family. The most recent clinical evidence search found no additional articles relating to the T-3® CT device family. The articles included a case report (Boullata et al.) and a prospective study (Scherer et al.). Bibliography: Scherer PF, Iizuka IJ, Ammirati AL, et al. Intermittent hemodiafiltration as a down-step transition therapy in patients with acute kidney injury admitted to intensive care unit who initially underwent continuous venovenous hemodiafiltration. The International Journal of Artificial Organs. 2021;44(4):223-228. Boullata JI, Stefankiewicz PM, Gallagher P, Gutsche J. Hemofilter Obstruction with Intravenous Lipid Administration. Journal of Parenteral and Enteral Nutrition. 2020;44(2):370-373.

• Source:PMCF_STHD_221

The dataset was provided by Scott O. Trerotola, MD an Interventional Radiologist at the Hospital of the University of Pennsylvania. Dr. Trerotola is also Stanley Baum Professor of Radiology, Professor of Radiology in Surgery, Vice Chair for Quality, Radiology, Associate Chair and Chief, Interventional Radiology, and Director, Penn HHT Center of Excellence at the Perelman School of Medicine at the University of Pennsylvania. The dataset includes information on 77 T-3® CT Catheter cases with dates of insertion ranging from 01 October 2009 to 14 September 2020 and dates of removal (or last known follow-up) ranging from 13 October 2009 to 23 September 2020. All insertion data is from the Interventional Radiology department at the University of Pennsylvania Hospital. 77 T-3® CT cases inclusive of several variant devices across catheter length (15cm, 20cm, 24cm) were collected. The following outcome measures were confirmed to be within State of the Art safety and performance outcome measures from published literature for Medcomp T-3® CT catheters.

  • Dwell Time (11.3 days 95%CI: 9.04 – 13.61 days)
  • Procedural Outcomes (100% 95%CI: 100% - 100%)
  • Catheter Related Blood Stream Infection (CRBSI) Rate (1.14 per 1,000 catheter days 95%CI: 0 – 6.39)
  • Catheter Associated Venous Thrombus (CAVT) Rate (0 per 1,000 catheter days 95%CI: 0 – 4.23)
  • Exit Site Infection (ESI) Rate (0 per 1,000 catheter days 95%CI: 0 – 4.23)
  • Power Injection Complications (0% 95%CI: 0% - 0%)
  • • Source:PMCF_Medcomp_211

    The Medcomp User Survey acquired responses from healthcare personnel familiar with any number of Medcomp’s product offerings. 20 respondents responded that they or their facility have used Medcomp short-term hemodialysis catheters, with 2 of those respondents using the T-3® CT device. There were no differences in mean user sentiments within short-term hemodialysis catheters across State of the Art Performance and Safety Outcome Measures or between device types relating to safety or performance. The following data points were collected from users of Medcomp short-term hemodialysis catheters (n=20):

  • (Mean Likert Scale Response) Catheters function as intended – 4.8 / 5
  • (Mean Likert Scale Response) Packaging allows for aseptic presentation – 4.9 / 5
  • (Mean Likert Scale Response) Benefit outweighs the risk – 4.7 / 5
  • Dwell Time (n=19) – 15.74 days (95%CI: 6.3 – 25.1) The following data points were collected from users of Medcomp T-3® CT catheters (n=2):
  • (Mean Likert Scale Response) Catheters function as intended – 5 / 5
  • (Mean Likert Scale Response) Packaging allows for aseptic presentation – 5 / 5
  • (Mean Likert Scale Response) Benefit outweighs the risk – 5 / 5
  • Dwell Time (n=2) – 7 days
  • • Source:Dr Trerotola Data Report

    The dataset was provided by Scott O. Trerotola, MD an Interventional Radiologist at the Hospital of the University of Pennsylvania. Dr. Trerotola is also Stanley Baum Professor of Radiology, Professor of Radiology in Surgery, Vice Chair for Quality, Radiology, Associate Chair and Chief, Interventional Radiology, and Director, Penn HHT Center of Excellence at the Perelman School of Medicine at the University of Pennsylvania. The dataset is consecutive, comprehensive, and includes catheter placements by interventional radiology Attending and Fellowship Physicians, as well as Residents under Attending supervision. This database was updated daily during the data collection period 02Jul2001-01Jul2019. 4 T-3® CT cases, including 24cm and 32cm length variants, were collected. The following outcome measures were collected for Medcomp T-3® CT devices:

  • Dwell Time – 3.8 days (Range: 1 – 9 days)
  • Procedural Outcomes – 100%
  • Power Injection Related Complications – No Events Reported
  • Catheter Related Blood Stream Infection – No Events Reported
  • Catheter Associated Venous Thrombus – No Events Reported
  • Exit Site Infection – No Events Reported
  • Source:T-3® CT Catheter Survey Report

    The T-3 Catheter Survey Report was intended to review and tabulate responses to ensure the device remains safe and effective when used as intended. Both survey reports concluded that the Medcomp® T-3® CT catheter is safe and effective when used as directed. One complication was reported in the first iteration of the survey, relating to bleeding. The second iteration of the survey indicated 3 of 4 respondents used the T-3® CT catheter for hemodialysis, 1 of 4 respondents used the T-3® CT catheter for apheresis, 4 of 4 respondents used the third internal lumen of the T-3® CT catheter for infusion therapy, and 3 of 4 respondents used the third internal lumen of the T-3® CT catheter for power injection of contrast media.

    Overall Summary of Clinical Safety and Performance

    T-3® CT Catheters facilitate hemodialysis, delivery of fluids and medications for treatments, and power injection of contrast media for CT examinations in patients in whom immediate central venous vascular access for short-term hemodialysis is deemed necessary based on the direction of a qualified, licensed physician. Upon review of the T-3® CT catheter data across all sources, it is possible to conclude that the benefits of the subject device outweigh the overall and individual risks when the device is used as intended by the manufacturer. It is the manufacturer’s and clinical expert evaluator’s opinion that activities both complete and ongoing are sufficient to support the safety, efficacy, and acceptable benefit/risk profile of the subject devices.

    Performance
    Outcome Benefit/Risk Acceptability Criteria Desired Trend Clinical Literature (Subject Device) PMCF Data (Subject Device)
    Performance Outcomes
    Dwell Time Greater than 8 days + ND*
    3.8 days (Range
    Procedural Outcomes Greater than 95% + ND*
    100% (Dr. Trerotola Data Report)
    Safety Outcomes
    Catheter Related Blood Stream Infection (CRBSI) Less than 7.8 incidents of CRBSI per 1,000 catheter days - No Events Reported (Summary of Published Literature)
    No Events Reported (Dr. Trerotola Data Report)
    Exit Site Infection Rate Less than 3.5 incidents of exit site infection per 1,000 catheter days - ND*
    No Events Reported (Dr. Trerotola Data Report)
    Catheter Associated Venous Thrombus (CAVT) Less than 11.4 incidents of CAVT per 1,000 catheter days - ND*
    No Events Reported (Dr. Trerotola Data Report)
    Power Injection Related Complications Less than 1.8% reported incidents of rupture and/or less than 15.4% reported incidents of displacement - ND*
    No Events Reported (Dr. Trerotola Data Report)
    *ND:indicates no data on the clinical data parameter **PMCF_Medcomp_211:asked respondents, if they agreed on a scale of 1 -5, that their experience in relation to each outcome was the same or better than the benefit/risk acceptability criteria.
    On-going or planned Post-Market Clinical Follow-up (PMCF)
    Outcome Benefit/Risk Acceptability Criteria Desired Trend Clinical Literature (Subject Device) PMCF Data (Subject Device)
    Performance Outcomes
    N/A N/A ?
    Safety Outcomes
    N/A N/A ?
    N/A:N/A

    On-going or Planned Post-Market Clinical Follow-Up (PMCF)

    Activity Description Reference Timeline
    Truveta Data Queries and Retrospective Analysis Collect additional clinical data on the device and comparators TBD Q4 2025
    State of the Art Literature Search Identify risks and trends with use of similar devices SAP-HD Q2 2025
    Clinical Evidence Literature Search Identify risks and trends with use of the device LRP-STHD Q3 2024
    Global Trial Database Search Identify ongoing clinical trials involving the devices N/A Q2 2025
    Multicenter Patient-Level Case Series Collect additional clinical data on the device PMCF_STHD_241 Q4 2025

    No emerging risks, complications or unexpected device failures have been detected from PMCF activities.

    6. Possible Therapeutic Alternatives

    The Kidney Disease Outcomes Quality Initiative (KDOQI) 2019 clinical practice guidelines have been used to support the below recommendations for treatments.

    Therapy Benefits Disadvantages Key Risks
    • AV Fistula
    • Permanent vascular access solution
    • Lower complication rate than hemodialysis via catheter.
    • Requires time to mature
    • Patients must sometimes self-cannulate
    • Stenosis
    • Thrombosis
    • Aneurysm
    • Pulmonary hypertension
    • Steal Syndrome
    • Septicemia
    • Hemodialysis Catheter
    • Useful for quick vascular access without AV Fistula in place
    • Can be used as a bridge dialysis method between other therapies
    • Not a permanent solution
    • Catheter dysfunction can disrupt regular treatment
    • Benefit is not equal for all patient populations
    • Post-procedural bleeding
    • Infection
    • Thrombosis
    • Decreased blood flow in dysfunctional catheter
    • Cardiovascular events
    • Fibrin sheath formation around catheter
    • Septicemia
    • Peritoneal Dialysis
    • Less restrictive diet than hemodialysis
    • Does not require hospitalization, can be done in any clean place
    • Clearance of impurities is limited by dialysate flow and peritoneal area
    • Peritonitis
    • Septicemia
    • Fluid overload
    • Kidney Transplant
    • Better quality of life compared to HD
    • Lower risk of death compared to HD
    • Fewer dietary restrictions compared to HD
    • Requires a donor which can take time
    • More risky for certain groups (aged, diabetics, etc.)
    • Patient must take rejection medication for life
    • Rejection medication has side effects
    • Thrombosis
    • Hemorrhage
    • Ureteral blockage
    • Infection
    • Organ rejection
    • Death
    • Myocardial infarction
    • Stroke
    • Comprehensive Conservative Care
    • Less imposed symptom burden than dialysis
    • Preserves life satisfaction
    • May aggravate clinical condition
    • Not designed to treat, but to minimize adverse events
    • Treatment may not actually minimize risks associated with CKD

    7. Suggested Profile and Training for Users

    The catheter should be inserted, manipulated, and removed by a qualified, licensed physician or other qualified health care professional under the direction of a physician.

    8. Reference to Any Harmonized Standards and Common Specifications (CS) Applied

    Harmonized Standard or CS Revision Title or Description Level of Compliance
    EN 556-1 2001 Sterilization of medical devices. Requirements for medical devices to be designated "STERILE". Requirements for terminally sterilized medical devices Full
    EN ISO 10555-1 2013+A1:2017 Intravascular catheters. Sterile and single-use catheters. General requirements Full
    EN ISO 10555-3 2013 Intravascular catheters. Sterile and single-use catheters. Central venous catheters Full
    EN ISO 10993-1 2020 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process Full
    EN ISO 10993-7 2008+ A1:2022 Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals — Amendment 1: Applicability of allowable limits for neonates and infants Full
    EN ISO 10993-18 2020 Biological evaluation of medical devices — Part 18: Chemical characterization of medical device materials within a risk management process Full
    EN ISO 11070 2014+A1:2018 Sterile single-use intravascular introducers, dilators and guidewires Full
    EN ISO 11135 2014 + A1: 2019 Sterilization of health-care products. Ethylene oxide. Requirements for the development, validation and routine control of a sterilization process for medical devices Full
    EN ISO 11138-1 2017 Sterilization of health care products — Biological indicators Part 1: General requirements Full
    EN ISO 11138-2 2017 Sterilization of health care products—Biological indicators—Part 2: Biological indicators for ethylene oxide sterilization processes Full
    EN ISO 11138-7 2019 Sterilization of health care products. Biological indicators - Guidance for the selection, use and interpretation of results Full
    EN ISO 11140-1 2014 Sterilization of health care products — Chemical indicators Part 1: General requirements Full
    EN ISO 11607-1 2020 Packaging for terminally sterilized medical devices. Requirements for materials, sterile barrier systems and packaging systems Full
    EN ISO 11607-2 2020 Packaging for terminally sterilized medical devices. Validation requirements for forming, sealing and assembly processes Full
    EN ISO 11737-1 2018 + A1: 2021 Sterilization of health care products. Microbiological methods. Determination of a population of microorganisms on products Full
    EN ISO 13485 2016 + A11: 2021 Medical Devices – Quality Management system – Requirements for Regulatory Purposes Full
    EN ISO 14155 2020 Clinical investigation of medical devices for human subjects — Good clinical practice Full
    EN ISO 14644-1 2015 Cleanrooms and associated controlled environments — Part 1: Classification of air cleanliness by particle concentration Full
    EN ISO 14644-2 2015 Cleanrooms and associated controlled environments — Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration Full
    EN ISO 14971 2019+A11:2021 Medical devices. Application of risk management to medical devices Full
    EN ISO 15223-1 2021 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements Full
    EN ISO/IEC 17025 2017 General requirements for the competence of testing and calibration laboratories Full
    PD CEN ISO/TR 20416 2020 Medical devices — post-market surveillance for manufacturers Full
    EN ISO 20417 2021 Medical devices - Information to be supplied by the manufacturer. Full
    EN 62366-1 2015 + A1: 2020 Medical devices — Part 1: Application of usability engineering to medical devices Full
    ISO 7000 2019 Graphical symbols for use on equipment. Registered symbols Partial
    ISO 594-1 1986 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 1: General requirements Full
    ISO 594-2 1998 Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment — Part 2: Lock Fittings Full
    MEDDEV 2.7.1 Rev 4 Clinical Evaluation: A Guide for Manufacturers and Notified Bodies Under Directives 93/42/EEC and 90/385/EEC Full
    MEDDEV 2.12/2 Rev. 2 GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP STUDIES A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES Full
    MDCG 2020-6 2020 Clinical evidence needed for medical devices previously CE marked under Directives 93/42/EEC or 90/385/EEC Full
    MDCG 2020-7 2020 Post-market clinical follow-up (PMCF) Plan Template A guide for manufacturers and notified bodies Full
    MDCG 2020-8 2020 Post-market clinical follow-up (PMCF) Evaluation Report Template A guide for manufacturers and notified bodies Full
    MDCG 2018-1 Rev. 4 Guidance on BASIC UDI-DI and changes to UDI-DI Full
    MDCG 2019-9 2022 Summary of safety and clinical performance Full
    ASTM D 4169-22 2022 Standard Practices for Performance Testing of Shipping Containers and Systems. Full
    ASTM F2096-11 2019 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) Full
    ASTM F2503-20 2020 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment Full
    ASTM F640-20 2020 Standard Test Methods for determining Radiopacity for Medical Use Full
    ASTM D4332-14 2014 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing Full
    Regulation (EU) 2017/745 2017 Regulation (EU) 2017/745 of the European Parliament and of the Council Full

    Revision History

    Revision Date CR# Author Description of Changes Validated
    1 07NOV2022 27445 KO Initial Implementation of SSCP No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    2 15MAR2023 27872 KO Updated SSCP in accordance with CER-026_C which includes the results of PMCF activity PMCF_STHD_221 and the addition of planned PMCF activity PMCF_STHD_241; updated language throughout the patient section to enhance readability No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    3 27APR2023 28089 GM Clinical benefit language has been updated according to the most recent revision of the instructions for use. No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device
    4 15APR2024 29025 GM Updated SSCP in accordance with CER-026_D which includes the addition of planned PMCF activity Truveta Data Queries and Retrospective Analysis and updated post market surveillance information No, this version was not validated by the Notified Body as this is a Class IIa or IIb implantable device

    Version 5.00 of Medical Components, Inc. Template QA-CL-200-1