Important Information
This Summary of Safety and Clinical Performance (SSCP) is intended
to provide public access to an updated summary of the main aspects
of the safety and clinical performance of the device. This SSCP is
not intended to replace the Instructions for Use as the main
document to ensure the safe use of the device, nor is it intended
to provide diagnostic or therapeutic suggestions to intended users
or patients.
Applicable Documents
| Document Type |
Document Title / Number |
| Design History File (DHF) |
09020 |
| ‘MDR Documentation’ File Number |
TD-026 |
1. Device Identification and General Information
Device Trade Name(s): T-3® CT Catheter
Manufacturer Name and Address: Medical
Components, Inc. (dba Medcomp®) 1499 Delp Drive Harleysville, PA
19438 USA
Manufacturer Single Registration Number (SRN):
US-MF-000008230
Basic UDI-DI: 00884908300MQ
Medical Device Nomenclature: F900201 – Temporary
Hemodialysis Catheters and Kits
Class of Device: III
Date First CE Certificate Issued: 11-Aug
Authorized Representative Name and SRN: Gerhard
Frömel European Regulatory Expert Medical Product Service GmbH
(MPS) Borngasse 20 35619 Braunfels, Germany SRN: US-MF-000008230
Notified Body Name and Single Identification Number:
BSI Netherlands NB2797
Device Grouping and Variants
The devices in scope of this document are all short-term
hemodialysis catheter sets. The device part numbers are organized
into variant categories. These devices are distributed as
procedure trays, in various configurations inclusive of
accessories and adjunctive devices (see section “Accessories
intended for use in combination with the Device”).
Variant Devices:
Variant Devices:
| Variant Description |
Part Number(s) |
Explanation of Multiple Part Numbers |
| 15.5F x 15cm Straight T-3 CT |
10468-815-001 |
|
| 15.5F x 20cm Straight T-3 CT |
10468-820-001 |
|
| 15.5F x 24cm Straight T-3 CT |
10468-824-001 |
|
| 15.5F x 28cm Straight T-3 CT |
10468-828-001 |
|
| 15.5F x 32cm Straight T-3 CT |
10468-832-001 |
|
Variant Devices:
| Variant Description |
Part Number(s) |
Explanation of Multiple Part Numbers |
| 15.5F x 15cm Straight T-3 CT |
10468-815-001 |
|
| 15.5F x 20cm Straight T-3 CT |
10468-820-001 |
|
| 15.5F x 24cm Straight T-3 CT |
10468-824-001 |
|
| 15.5F x 28cm Straight T-3 CT |
10468-828-001 |
|
| 15.5F x 32cm Straight T-3 CT |
10468-832-001 |
|
Procedure Trays:
Procedure Trays:
| Catalog Code |
Part Number |
Description |
| MC013501 |
10468-815-001 |
15.5F X 15cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013502 |
10468-820-001 |
15.5F X 20cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013503 |
10468-824-001 |
15.5F X 24cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013504 |
10468-828-001 |
15.5F X 28cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013505 |
10468-832-001 |
15.5F X 32cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
Procedure Trays:
| Catalog Code |
Part Number |
Description |
| MC013501 |
10468-815-001 |
15.5F X 15cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013502 |
10468-820-001 |
15.5F X 20cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013503 |
10468-824-001 |
15.5F X 24cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013504 |
10468-828-001 |
15.5F X 28cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
| MC013505 |
10468-832-001 |
15.5F X 32cm T-3 CT Power Injectable Triple Lumen
Hemodialysis Catheter Basic Set
|
Configurations of Procedure Trays:
| Configuration Type |
Kit Components |
| T-3® CT Basic Set |
(1) Catheter (1) Guidewire (1) Guidewire Advancer (1) Needle
(1) Scalpel (1) Suture (1) Peelable Introducer (2) Dilator
(3) End Cap
|
2. Intended Use of the Device
Intended Purpose: T-3® CT Catheters are intended
for use in adult patients with Acute Kidney Injury (AKI) or
Chronic Kidney Disease (CKD) for whom immediate central venous
vascular access for short-term hemodialysis, intravenous
administration of fluids or medications, and power injection of
contrast media is deemed necessary based on the direction of a
qualified, licensed physician. The catheter is intended to be used
under the regular review and assessment of qualified health
professionals. This catheter is for Single Use Only.
Indication(s): The T-3® CT Catheter is indicated
for short-term use where vascular access is required for less than
14 days for the purpose of hemodialysis. The third internal lumen
is indicated for intravenous administration of fluids or
medications and power injection of contrast media.
Target Population(s): T-3® CT Catheters are
intended for use in adult patients with Acute Kidney Injury (AKI)
or Chronic Kidney Disease (CKD) for whom immediate central venous
vascular access for short-term hemodialysis, intravenous
administration of fluids or medications, and power injection of
contrast media is deemed necessary based on the direction of a
qualified, licensed physician. The catheter is not intended for
use in pediatric patients.
Contraindications and/or Limitations:
-
Known or suspected allergies to any of the components of the
catheter or the kit.
-
This device is contraindicated for patients exhibiting severe,
uncontrolled coagulopathy or thrombocytopenia.
3. Device Description
Device Name: T-3® CT Catheter
Description of Device: T-3® CT Catheter:The T-3®
Catheter is a short-term dialysis catheter made of thermosensitive
polyurethane. The catheter has three separate lumens allowing
continuous blood flow. The venous (blue) and arterial (red) lumens
may be used for hemodialysis treatments. The middle (purple) lumen
is independent from the two dialysis lumens, and may be used for
administration of fluids or medications and power injection of
contrast media. The catheter is available in a variety of lengths
to accommodate physician preference and clinical needs.
Device Name: T-3® CT Catheter
Description of Device: T-3® CT Catheter:The T-3®
Catheter is a short-term dialysis catheter made of thermosensitive
polyurethane. The catheter has three separate lumens allowing
continuous blood flow. The venous (blue) and arterial (red) lumens
may be used for hemodialysis treatments. The middle (purple) lumen
is independent from the two dialysis lumens, and may be used for
administration of fluids or medications and power injection of
contrast media. The catheter is available in a variety of lengths
to accommodate physician preference and clinical needs.
Materials / Substances in Contact with Patient Tissue:
The percentage ranges in the table below are based on the weights
of the 15cm catheter (15.15g) and the 32cm catheter (17.91g).
T-3® CT Catheter
| Material |
% Weight (w/w) |
| Polyurethane |
64.09 - 66.55 |
| Acetal co-polymer |
16.46 - 19.46 |
| Acrylonitrile Butadiene Styrene |
5.57 - 6.59 |
| Silicone |
5.57 - 6.59 |
| Barium sulfate |
3.28 - 5.86 |
The percentage ranges in the table below are based on the weights
of the 15cm catheter (15.15g) and the 32cm catheter (17.91g).
T-3® CT Catheter
| Material |
% Weight (w/w) |
| Polyurethane |
64.09 - 66.55 |
| Acetal co-polymer |
16.46 - 19.46 |
| Acrylonitrile Butadiene Styrene |
5.57 - 6.59 |
| Silicone |
5.57 - 6.59 |
| Barium sulfate |
3.28 - 5.86 |
Note:Per the instructions for use, the device is contraindicated
for patients with known or suspected allergies to the above
materials.
Information on Medicinal Substances in the Device:
How the Device Achieves its Intended Mode of Action:
Hemodialysis catheters are centrally placed access tubes. A
typical hemodialysis catheter uses a thin, flexible tube. The tube
has two openings. The tube goes into a large vein. The vein is
usually the internal jugular vein. Blood withdraws through one
lumen of the catheter. The blood flows to the dialysis machine
through a separate tubing set. The blood is then processed and
filtered. The blood returns to the patient through the second
lumen. This device is used when dialysis must start at once.
Patients may not have a functioning AV fistula or graft. Catheter
hemodialysis normally happens on a short-term basis.
Sterilization Information: Contents sterile and
non-pyrogenic in unopened, undamaged package. Sterilized by
Ethylene Oxide.
Previous Generations / Variants:
| Name of Previous Generation |
Differences from Current Device |
| N/A |
N/A |
Accessories Intended for Use in Combination with the
Device:
| Name of Accessory |
Description of Accessory |
| Guidewire |
For general intravascular use to facilitate the selective
placement of medical devices in the vessel anatomy.
|
| Guidewire Advancer |
Aid for introduction of guidewire into target vein.
|
| Introducer Needle |
Used for the percutaneous introduction of guidewires.
|
| Dilator |
Designed for percutaneous entry into a vessel in order to
enlarge the opening of the vessel for the placement of a
catheter in a vein.
|
| Valved Peelable Introducer |
Intended to obtain central venous access to facilitate
catheter insertion into the central venous system.
|
| Scalpel |
A cutting device during surgical, pathology and minor
medical procedures
|
| Suture |
Used to secure the catheter to the body.
|
| End Cap |
To keep clean and protect catheter luer between treatments.
|
| Syringe |
Attached to introducer needle to help capture blood return
once introducer needle perforates targeted vein, prevent air
embolism
|
| Adhesive Wound Dressing |
Intended to protect the catheter from contamination when not
in use
|
Other Devices or Products Intended for Use in Combination with
the Device:
| Name of Device or Product |
Description of Device or Product |
| N/A |
N/A |
4. Risks and Warnings
Residual Risks and Undesirable Effects: All
surgical procedures carry risk. Medcomp® has implemented risk
management processes to proactively find and mitigate these risks
as far as possible without adversely affecting the benefit-risk
profile of the device. After mitigation, residual risks and the
possibility of adverse events from use of this product remain.
Medcomp® has determined that all residual risks are acceptable
when considered with respect to the expected clinical benefits of
the T-3® CT Catheter and the benefits of other similar
hemodialysis devices.
| Residual Harm Type |
Possible Adverse Events Associated with Harm
|
| Allergic Reaction |
Allergic Reaction
|
| Bleeding |
Bleeding (May be severe)
|
| Cardiac Event |
Cardiac Arrhythmia
|
| Embolism |
Air Embolus
|
| Infection |
Bacteremia
|
| Perforation |
Inferior Vena Cava Puncture
|
| Stenosis |
Venous Stenosis
|
| Tissue Injury |
Brachial Plexus Injury
|
| Thrombosis |
Central Venous Thrombosis
|
| Miscellaneous Complications |
Catheter Dysfunction
|
|
Quantification of Residual Risks
|
|
PMS Complaints 01 January 2018 – 31 December 2023
|
PMCF Events |
|
Units Sold: 21,463 |
Units Studied: 81 |
|
Patient Residual Harm Category
|
% of Devices |
% of Devices |
| Allergic Reaction |
Not Reported |
Not Reported |
| Bleeding |
Not Reported |
Not Reported |
| Cardiac Event |
Not Reported |
Not Reported |
| Embolism |
Not Reported |
Not Reported |
| Infection |
Not Reported |
1.23% |
| Perforation |
Not Reported |
Not Reported |
| Stenosis |
Not Reported |
Not Reported |
| Tissue Injury |
Not Reported |
Not Reported |
| Thrombosis |
Not Reported |
2.47% |
Warnings and Precautions:
Warnings listed for the T-3® CT Catheter are as follows:
-
Do not insert catheter in thrombosed vessels.
-
Do not advance the guidewire or catheter if unusual resistance
is encountered.
-
Do not insert or withdraw the guidewire forcibly from any
component. If the guidewire becomes damaged, guidewire and any
associated componentry must be removed together.
-
Do not resterilize the catheter or accessories by any method.
-
Contents sterile and non-pyrogenic in unopened, undamaged
package. STERILIZED BY ETHYLENE OXIDE
-
Do not re-use catheter or accessories as there may be a failure
to adequately clean and decontaminate the device which may lead
to contamination, catheter degradation, device fatigue, or
endotoxin reaction.
-
Do not use catheter or accessories if package is opened or
damaged.
-
Do not use catheter or accessories if any sign of product damage
is visible or the use-by date has passed.
-
Do not use sharp instruments near the extension tubing or
catheter lumen.
-
Do not use scissors to remove dressing. Precautions listed for
T-3® CT Catheter are as follows:
-
Examine catheter lumen and extensions before and after each
treatment for damage.
-
To prevent accidents, ensure the security of all caps and
bloodline connections prior to and
- between treatments.
-
Use only Luer Lock (threaded) Connectors with this catheter.
-
In the rare event that a hub or connector separates from any
component during insertion or use, take all necessary steps and
precautions to prevent blood loss or air embolism and remove the
catheter.
-
Before attempting catheter insertion, ensure that you are
familiar with the potential complications and their emergency
treatment should any of them occur.
-
Repeated overtightening of bloodlines, syringes, and caps will
reduce connector life and could lead to potential connector
failure.
-
The catheter will be damaged if clamps other than what is
provided with this kit are used.
-
Avoid clamping near the Luer Lock and hub of the catheter.
Clamping of the tubing repeatedly in the same location may
weaken tubing.
-
The third lumen of the catheter allows for intravenous
administration of fluids or medications and power injection of
contrast media. Refer to standards of practice and institutional
policies for compatible infusion agents for central venous
access.
-
Follow all contraindications, warnings, precautions, and
instructions for all infusates including contrast media as
specified by their manufacturer.
-
The red arterial and blue venous lumens should not be used for
infusion of any infusates or contrast media as patient injury
may occur.
-
The center (distal) infusion lumen should not be used for
hemodialysis as insufficient treatment may occur.
Other Relevant Aspects of Safety: For a period of
01 January 2019 to 31 December 2023 there was 1 complaint for
9,842 units sold, giving an overall complaint rate of 0.01%. There
were no death-related events. No events resulted in recalls during
the review period.
5. Summary of Clinical Evaluation and Post-Market Clinical
Follow-Up (PMCF)
Summary of Clinical Data Related to Subject Device
Specific Case Numbers (Mixed Cohort Case Numbers) Identified and
Used for Clinical Performance Evaluation
| Product Family |
Clinical Literature |
PMCF Data |
Total Cases |
User Survey Responses |
| Clinical Literature |
1 (& 10 Mixed Cohort Cases) |
|
|
|
| PMCF Data |
81 |
|
|
|
| Total Cases |
82 (& 10 Mixed Cohort Cases) |
|
|
|
| User Survey Responses |
19 |
|
|
|
Clinical performance was measured using parameters including but
not limited to dwell time, catheter insertion outcomes, and
adverse event rates. Critical clinical parameters extracted from
these studies met standards set forth in the guidelines for the
State of the Art. There were no unforeseen adverse events or other
high occurrences of adverse events detected in any of the clinical
activities. Medcomp® STHD catheters are subjected to, and must
pass, simulated use testing intended to replicate 30 days use as
part of device development. The T-3® CT Catheter passed this
testing. Clinical guidelines recommend to limit the use of
temporary, noncuffed, nontunneled dialysis catheters to a maximum
of 2 weeks (KDOQI 2019), however, duration of use of these
catheters has varied in available clinical evidence identified by
the manufacturer to date. Although Medcomp® catheters materials
contain non-degradable polymers, fully functional catheters may be
removed for other reasons, such as intractable infection or change
of therapy. Published clinical literature does not always focus on
the physical lifetime of a catheter for these reasons. In the case
of the T-3® Catheter, 77 catheters had a mean dwell time of 11.3
days [95%CI: 9.04 – 13.61 days] duration of use that has been
found in clinical use reported to date. Based on this information
The T-3® CT Catheter has a 30 day lifetime; however, the decision
to remove and/or replace the catheter should be based on clinical
performance and need, and not any predetermined point in time.
Summary of Clinical Data Related to the Equivalent Device
Clinical evidence from published literature and PMCF activities
has been generated specific to known and unknown variants of the
subject device. The equivalency rationale in the updated clinical
evaluation report will demonstrate that the clinical evidence
available for these variants is representative of the range of
device variants in the device family. There are no clinical or
biological differences between variants within the subject device
family, and the potential impact of the technical differences will
be rationalized in the updated clinical evaluation report.
Summary of Clinical Data from Pre-Market Investigations (if
applicable)
No pre-market clinical devices were used for the device’s clinical
evaluation.
Summary of Clinical Data from Other Sources
Source:Summary of Published Literature
Previous clinical evidence literature searches found two published
literature articles representing 1 T-3® CT device family specific
case, and an additional 10 mixed cohort cases inclusive of the
T-3® CT device family. The most recent clinical evidence search
found no additional articles relating to the T-3® CT device
family. The articles included a case report (Boullata et al.) and
a prospective study (Scherer et al.). Bibliography: Scherer PF,
Iizuka IJ, Ammirati AL, et al. Intermittent hemodiafiltration as a
down-step transition therapy in patients with acute kidney injury
admitted to intensive care unit who initially underwent continuous
venovenous hemodiafiltration. The International Journal of
Artificial Organs. 2021;44(4):223-228. Boullata JI, Stefankiewicz
PM, Gallagher P, Gutsche J. Hemofilter Obstruction with
Intravenous Lipid Administration. Journal of Parenteral and
Enteral Nutrition. 2020;44(2):370-373.
• Source:PMCF_STHD_221
The dataset was provided by Scott O. Trerotola, MD an
Interventional Radiologist at the Hospital of the University of
Pennsylvania. Dr. Trerotola is also Stanley Baum Professor of
Radiology, Professor of Radiology in Surgery, Vice Chair for
Quality, Radiology, Associate Chair and Chief, Interventional
Radiology, and Director, Penn HHT Center of Excellence at the
Perelman School of Medicine at the University of Pennsylvania. The
dataset includes information on 77 T-3® CT Catheter cases with
dates of insertion ranging from 01 October 2009 to 14 September
2020 and dates of removal (or last known follow-up) ranging from
13 October 2009 to 23 September 2020. All insertion data is from
the Interventional Radiology department at the University of
Pennsylvania Hospital. 77 T-3® CT cases inclusive of several
variant devices across catheter length (15cm, 20cm, 24cm) were
collected. The following outcome measures were confirmed to be
within State of the Art safety and performance outcome measures
from published literature for Medcomp T-3® CT catheters.
Dwell Time (11.3 days 95%CI: 9.04 – 13.61 days)
Procedural Outcomes (100% 95%CI: 100% - 100%)
Catheter Related Blood Stream Infection (CRBSI) Rate (1.14 per
1,000 catheter days 95%CI: 0 – 6.39)
Catheter Associated Venous Thrombus (CAVT) Rate (0 per 1,000
catheter days 95%CI: 0 – 4.23)
Exit Site Infection (ESI) Rate (0 per 1,000 catheter days 95%CI: 0
– 4.23)
Power Injection Complications (0% 95%CI: 0% - 0%)
• Source:PMCF_Medcomp_211
The Medcomp User Survey acquired responses from healthcare
personnel familiar with any number of Medcomp’s product offerings.
20 respondents responded that they or their facility have used
Medcomp short-term hemodialysis catheters, with 2 of those
respondents using the T-3® CT device. There were no differences in
mean user sentiments within short-term hemodialysis catheters
across State of the Art Performance and Safety Outcome Measures or
between device types relating to safety or performance. The
following data points were collected from users of Medcomp
short-term hemodialysis catheters (n=20):
(Mean Likert Scale Response) Catheters function as intended – 4.8
/ 5
(Mean Likert Scale Response) Packaging allows for aseptic
presentation – 4.9 / 5
(Mean Likert Scale Response) Benefit outweighs the risk – 4.7 / 5
Dwell Time (n=19) – 15.74 days (95%CI: 6.3 – 25.1) The following
data points were collected from users of Medcomp T-3® CT catheters
(n=2):
(Mean Likert Scale Response) Catheters function as intended – 5 /
5
(Mean Likert Scale Response) Packaging allows for aseptic
presentation – 5 / 5
(Mean Likert Scale Response) Benefit outweighs the risk – 5 / 5
Dwell Time (n=2) – 7 days
• Source:Dr Trerotola Data Report
The dataset was provided by Scott O. Trerotola, MD an
Interventional Radiologist at the Hospital of the University of
Pennsylvania. Dr. Trerotola is also Stanley Baum Professor of
Radiology, Professor of Radiology in Surgery, Vice Chair for
Quality, Radiology, Associate Chair and Chief, Interventional
Radiology, and Director, Penn HHT Center of Excellence at the
Perelman School of Medicine at the University of Pennsylvania. The
dataset is consecutive, comprehensive, and includes catheter
placements by interventional radiology Attending and Fellowship
Physicians, as well as Residents under Attending supervision. This
database was updated daily during the data collection period
02Jul2001-01Jul2019. 4 T-3® CT cases, including 24cm and 32cm
length variants, were collected. The following outcome measures
were collected for Medcomp T-3® CT devices:
Dwell Time – 3.8 days (Range: 1 – 9 days)
Procedural Outcomes – 100%
Power Injection Related Complications – No Events Reported
Catheter Related Blood Stream Infection – No Events Reported
Catheter Associated Venous Thrombus – No Events Reported
Exit Site Infection – No Events Reported
Source:T-3® CT Catheter Survey Report
The T-3 Catheter Survey Report was intended to review and tabulate
responses to ensure the device remains safe and effective when
used as intended. Both survey reports concluded that the Medcomp®
T-3® CT catheter is safe and effective when used as directed. One
complication was reported in the first iteration of the survey,
relating to bleeding. The second iteration of the survey indicated
3 of 4 respondents used the T-3® CT catheter for hemodialysis, 1
of 4 respondents used the T-3® CT catheter for apheresis, 4 of 4
respondents used the third internal lumen of the T-3® CT catheter
for infusion therapy, and 3 of 4 respondents used the third
internal lumen of the T-3® CT catheter for power injection of
contrast media.
Overall Summary of Clinical Safety and Performance
T-3® CT Catheters facilitate hemodialysis, delivery of fluids and
medications for treatments, and power injection of contrast media
for CT examinations in patients in whom immediate central venous
vascular access for short-term hemodialysis is deemed necessary
based on the direction of a qualified, licensed physician. Upon
review of the T-3® CT catheter data across all sources, it is
possible to conclude that the benefits of the subject device
outweigh the overall and individual risks when the device is used
as intended by the manufacturer. It is the manufacturer’s and
clinical expert evaluator’s opinion that activities both complete
and ongoing are sufficient to support the safety, efficacy, and
acceptable benefit/risk profile of the subject devices.
Performance
| Outcome |
Benefit/Risk Acceptability Criteria |
Desired Trend |
Clinical Literature (Subject Device) |
PMCF Data (Subject Device) |
| Performance Outcomes |
| Dwell Time |
Greater than 8 days |
+
|
ND*
|
3.8 days (Range
|
| Procedural Outcomes |
Greater than 95% |
+
|
ND*
|
100% (Dr. Trerotola Data Report)
|
| Safety Outcomes |
|
Catheter Related Blood Stream Infection (CRBSI)
|
Less than 7.8 incidents of CRBSI per 1,000 catheter days
|
-
|
No Events Reported (Summary of Published Literature)
|
No Events Reported (Dr. Trerotola Data Report)
|
| Exit Site Infection Rate |
Less than 3.5 incidents of exit site infection per 1,000
catheter days
|
-
|
ND*
|
No Events Reported (Dr. Trerotola Data Report)
|
|
Catheter Associated Venous Thrombus (CAVT)
|
Less than 11.4 incidents of CAVT per 1,000 catheter days
|
-
|
ND*
|
No Events Reported (Dr. Trerotola Data Report)
|
| Power Injection Related Complications |
Less than 1.8% reported incidents of rupture and/or less
than 15.4% reported incidents of displacement
|
-
|
ND*
|
No Events Reported (Dr. Trerotola Data Report)
|
*ND:indicates no data on the clinical data parameter
**PMCF_Medcomp_211:asked respondents, if they agreed on a scale
of 1 -5, that their experience in relation to each outcome was the
same or better than the benefit/risk acceptability criteria.
On-going or planned Post-Market Clinical Follow-up (PMCF)
| Outcome |
Benefit/Risk Acceptability Criteria |
Desired Trend |
Clinical Literature (Subject Device) |
PMCF Data (Subject Device) |
| Performance Outcomes |
| N/A |
N/A |
?
|
|
|
| Safety Outcomes |
| N/A |
N/A |
?
|
|
|
N/A:N/A
On-going or Planned Post-Market Clinical Follow-Up (PMCF)
| Activity |
Description |
Reference |
Timeline |
|
Truveta Data Queries and Retrospective Analysis
|
Collect additional clinical data on the device and
comparators
|
TBD |
Q4 2025 |
| State of the Art Literature Search |
Identify risks and trends with use of similar devices
|
SAP-HD |
Q2 2025 |
| Clinical Evidence Literature Search |
Identify risks and trends with use of the device
|
LRP-STHD |
Q3 2024 |
| Global Trial Database Search |
Identify ongoing clinical trials involving the devices
|
N/A |
Q2 2025 |
| Multicenter Patient-Level Case Series |
Collect additional clinical data on the device
|
PMCF_STHD_241 |
Q4 2025 |
No emerging risks, complications or unexpected device failures
have been detected from PMCF activities.
6. Possible Therapeutic Alternatives
The Kidney Disease Outcomes Quality Initiative (KDOQI) 2019
clinical practice guidelines have been used to support the below
recommendations for treatments.
| Therapy |
Benefits |
Disadvantages |
Key Risks |
| • AV Fistula |
-
Permanent vascular access solution
-
Lower complication rate than hemodialysis via catheter.
|
- Requires time to mature
-
Patients must sometimes self-cannulate
|
- Stenosis
- Thrombosis
- Aneurysm
- Pulmonary hypertension
- Steal Syndrome
- Septicemia
|
| • Hemodialysis Catheter |
-
Useful for quick vascular access without AV Fistula in
place
-
Can be used as a bridge dialysis method between other
therapies
|
- Not a permanent solution
-
Catheter dysfunction can disrupt regular treatment
-
Benefit is not equal for all patient populations
|
- Post-procedural bleeding
- Infection
- Thrombosis
-
Decreased blood flow in dysfunctional catheter
- Cardiovascular events
-
Fibrin sheath formation around catheter
- Septicemia
|
| • Peritoneal Dialysis |
-
Less restrictive diet than hemodialysis
-
Does not require hospitalization, can be done in any
clean place
|
-
Clearance of impurities is limited by dialysate flow and
peritoneal area
|
- Peritonitis
- Septicemia
- Fluid overload
|
| • Kidney Transplant |
-
Better quality of life compared to HD
-
Lower risk of death compared to HD
-
Fewer dietary restrictions compared to HD
|
-
Requires a donor which can take time
-
More risky for certain groups (aged, diabetics, etc.)
-
Patient must take rejection medication for life
-
Rejection medication has side effects
|
- Thrombosis
- Hemorrhage
- Ureteral blockage
- Infection
- Organ rejection
- Death
- Myocardial infarction
- Stroke
|
| • Comprehensive Conservative Care |
-
Less imposed symptom burden than dialysis
- Preserves life satisfaction
|
- May aggravate clinical condition
-
Not designed to treat, but to minimize adverse events
|
-
Treatment may not actually minimize risks associated
with CKD
|
7. Suggested Profile and Training for Users
The catheter should be inserted, manipulated, and removed by a
qualified, licensed physician or other qualified health care
professional under the direction of a physician.
8. Reference to Any Harmonized Standards and Common Specifications
(CS) Applied
| Harmonized Standard or CS |
Revision |
Title or Description |
Level of Compliance |
| EN 556-1 |
2001 |
Sterilization of medical devices. Requirements for medical
devices to be designated "STERILE". Requirements for
terminally sterilized medical devices
|
Full |
| EN ISO 10555-1 |
2013+A1:2017 |
Intravascular catheters. Sterile and single-use catheters.
General requirements
|
Full |
| EN ISO 10555-3 |
2013 |
Intravascular catheters. Sterile and single-use catheters.
Central venous catheters
|
Full |
| EN ISO 10993-1 |
2020 |
Biological evaluation of medical devices — Part 1:
Evaluation and testing within a risk management process
|
Full |
| EN ISO 10993-7 |
2008+ A1:2022 |
Biological evaluation of medical devices — Part 7: Ethylene
oxide sterilization residuals — Amendment 1: Applicability
of allowable limits for neonates and infants
|
Full |
| EN ISO 10993-18 |
2020 |
Biological evaluation of medical devices — Part 18: Chemical
characterization of medical device materials within a risk
management process
|
Full |
| EN ISO 11070 |
2014+A1:2018 |
Sterile single-use intravascular introducers, dilators and
guidewires
|
Full |
| EN ISO 11135 |
2014 + A1: 2019 |
Sterilization of health-care products. Ethylene oxide.
Requirements for the development, validation and routine
control of a sterilization process for medical devices
|
Full |
| EN ISO 11138-1 |
2017 |
Sterilization of health care products — Biological
indicators Part 1: General requirements
|
Full |
| EN ISO 11138-2 |
2017 |
Sterilization of health care products—Biological
indicators—Part 2: Biological indicators for ethylene oxide
sterilization processes
|
Full |
| EN ISO 11138-7 |
2019 |
Sterilization of health care products. Biological indicators
- Guidance for the selection, use and interpretation of
results
|
Full |
| EN ISO 11140-1 |
2014 |
Sterilization of health care products — Chemical indicators
Part 1: General requirements
|
Full |
| EN ISO 11607-1 |
2020 |
Packaging for terminally sterilized medical devices.
Requirements for materials, sterile barrier systems and
packaging systems
|
Full |
| EN ISO 11607-2 |
2020 |
Packaging for terminally sterilized medical devices.
Validation requirements for forming, sealing and assembly
processes
|
Full |
| EN ISO 11737-1 |
2018 + A1: 2021 |
Sterilization of health care products. Microbiological
methods. Determination of a population of microorganisms on
products
|
Full |
| EN ISO 13485 |
2016 + A11: 2021 |
Medical Devices – Quality Management system – Requirements
for Regulatory Purposes
|
Full |
| EN ISO 14155 |
2020 |
Clinical investigation of medical devices for human subjects
— Good clinical practice
|
Full |
| EN ISO 14644-1 |
2015 |
Cleanrooms and associated controlled environments — Part 1:
Classification of air cleanliness by particle concentration
|
Full |
| EN ISO 14644-2 |
2015 |
Cleanrooms and associated controlled environments — Part 2:
Monitoring to provide evidence of cleanroom performance
related to air cleanliness by particle concentration
|
Full |
| EN ISO 14971 |
2019+A11:2021 |
Medical devices. Application of risk management to medical
devices
|
Full |
| EN ISO 15223-1 |
2021 |
Medical devices — Symbols to be used with medical device
labels, labelling and information to be supplied — Part 1:
General requirements
|
Full |
| EN ISO/IEC 17025 |
2017 |
General requirements for the competence of testing and
calibration laboratories
|
Full |
| PD CEN ISO/TR 20416 |
2020 |
Medical devices — post-market surveillance for manufacturers
|
Full |
| EN ISO 20417 |
2021 |
Medical devices - Information to be supplied by the
manufacturer.
|
Full |
| EN 62366-1 |
2015 + A1: 2020 |
Medical devices — Part 1: Application of usability
engineering to medical devices
|
Full |
| ISO 7000 |
2019 |
Graphical symbols for use on equipment. Registered symbols
|
Partial |
| ISO 594-1 |
1986 |
Conical fittings with a 6 % (Luer) taper for syringes,
needles and certain other medical equipment — Part 1:
General requirements
|
Full |
| ISO 594-2 |
1998 |
Conical fittings with a 6 % (Luer) taper for syringes,
needles and certain other medical equipment — Part 2: Lock
Fittings
|
Full |
| MEDDEV 2.7.1 |
Rev 4 |
Clinical Evaluation: A Guide for Manufacturers and Notified
Bodies Under Directives 93/42/EEC and 90/385/EEC
|
Full |
| MEDDEV 2.12/2 |
Rev. 2 |
GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP
STUDIES A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES
|
Full |
| MDCG 2020-6 |
2020 |
Clinical evidence needed for medical devices previously CE
marked under Directives 93/42/EEC or 90/385/EEC
|
Full |
| MDCG 2020-7 |
2020 |
Post-market clinical follow-up (PMCF) Plan Template A guide
for manufacturers and notified bodies
|
Full |
| MDCG 2020-8 |
2020 |
Post-market clinical follow-up (PMCF) Evaluation Report
Template A guide for manufacturers and notified bodies
|
Full |
| MDCG 2018-1 |
Rev. 4 |
Guidance on BASIC UDI-DI and changes to UDI-DI
|
Full |
| MDCG 2019-9 |
2022 |
Summary of safety and clinical performance
|
Full |
| ASTM D 4169-22 |
2022 |
Standard Practices for Performance Testing of Shipping
Containers and Systems.
|
Full |
| ASTM F2096-11 |
2019 |
Standard Test Method for Detecting Gross Leaks in Packaging
by Internal Pressurization (Bubble Test)
|
Full |
| ASTM F2503-20 |
2020 |
Standard Practice for Marking Medical Devices and Other
Items for Safety in the Magnetic Resonance Environment
|
Full |
| ASTM F640-20 |
2020 |
Standard Test Methods for determining Radiopacity for
Medical Use
|
Full |
| ASTM D4332-14 |
2014 |
Standard Practice for Conditioning Containers, Packages, or
Packaging Components for Testing
|
Full |
| Regulation (EU) 2017/745 |
2017 |
Regulation (EU) 2017/745 of the European Parliament and of
the Council
|
Full |
Revision History
| Revision |
Date |
CR# |
Author |
Description of Changes |
Validated |
| 1 |
07NOV2022 |
27445 |
KO |
Initial Implementation of SSCP |
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 2 |
15MAR2023 |
27872 |
KO |
Updated SSCP in accordance with CER-026_C which includes the
results of PMCF activity PMCF_STHD_221 and the addition of
planned PMCF activity PMCF_STHD_241; updated language
throughout the patient section to enhance readability
|
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 3 |
27APR2023 |
28089 |
GM |
Clinical benefit language has been updated according to the
most recent revision of the instructions for use.
|
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|
| 4 |
15APR2024 |
29025 |
GM |
Updated SSCP in accordance with CER-026_D which includes the
addition of planned PMCF activity Truveta Data Queries and
Retrospective Analysis and updated post market surveillance
information
|
No, this version was not validated by the Notified Body as
this is a Class IIa or IIb implantable device
|